Primary stability of a cementless modular revision hip stem in relation with the femoral defect size: A biomechanical study

Affiliations
Free article

Primary stability of a cementless modular revision hip stem in relation with the femoral defect size: A biomechanical study

Martin Ellenrieder et al. J Orthop Surg (Hong Kong). 2020 Sep-Dec.
Free article

Abstract

Purpose: Cementless modular fluted hip stems are commonly used in revision arthroplasty. Nevertheless, there is a wide spectrum of recommendations concerning the minimum bone stock required to enable osseous ingrowth and implant-bone micromotions <100 µm. This experimental study investigated the primary stability of a tapered cementless fluted revision stem depending on different types of bone defects.

Methods: Implant-bone interface movements with a bimodular stem were examined under cyclic axial and torsional loading using composite femora. In four degrees of freedom, the implant subsidence and micromotions were captured with linear variable differential transformers for the intact femora and seven different defects ranging from Paprosky type I to type IIIB.

Results: With a 7-cm length of intact diaphysis proximal to the isthmus (Paprosky IIIA), mean implant-bone micromotions of 66 µm occurred. An implant-bone contact zone of only 5 cm (Paprosky IIIA) resulted in micromotions notably over 100 µm and significantly increased subsidence (p < 0.05). With a Paprosky IIIB defect (3 cm of intact diaphysis) rotational instability occurred in all specimens.

Conclusions: Aside from critically increased interfacial micromotions (>100 µm), rotational instability emerged as a mechanism of fixation failure when the implant-bone contact zone was only 5 cm or less. Hence, future studies investigating the implant fixation in the case of femoral bone defects should consider both axial and torsional loading. With regard to the clinical application, our data suggest maintaining 7 cm of diaphyseal implant-bone contact for a safe anchorage of cementless fluted hip revision stems.

Keywords: cementless; fluted; hip arthroplasty; micromotion; primary stability; revision stem; subsidence.