Ultrasound-guided intraoperative electron beam radiation therapy: A phantom study
- PMID: 32911372
- DOI: 10.1016/j.ejmp.2020.06.021
Ultrasound-guided intraoperative electron beam radiation therapy: A phantom study
Abstract
Purpose: To develop the method for ultrasound (US)-guided intra-operative electron beam radiation therapy (IOERT).
Methods: We first established the simulation, planning, and delivery methods for US-guided IOERT and constructed appropriate hardware (the multi-function applicator, accessories, and US phantom). We tested our US-guided IOERT method using this hardware and the Monte Carlo simulation IOERT treatment planning system (TPS). The IOERT TPS used a compensator to build the conformal dose distribution. Then, we used the TPS to evaluate the effect of setup uncertainty on target coverage by introducing phantom setup error ranging from 0 mm to 10 mm to the plans with and without the compensator.
Results: The simulation, planning, and delivery methods for US-guided IOERT were introduced and validated on a phantom. A complete technique for US-guided IOERT was established. Target coverage decreased by about 12% and 29% as the phantom setup error increased to 5 mm and 10 mm for the plans with compensator, respectively. Without compensator, the corresponding target coverage decreases were 2% and 13%, respectively.
Conclusion: In our study, we developed the multi-function applicator, US Phantom, and TPS for IOERT. The procedures included not only dose distribution planning, but also intraoperative US imaging, which provided the information necessary during surgery to improve IOERT quality assurance. Target coverage was more sensitive to setup errors with compensator compared to no compensator. Further studies are needed to validate the clinical efficacy of this US-guided IOERT method.
Keywords: Image-guided radiation therapy; Intraoperative electron beam radiation therapy; Ultrasound imaging.
Copyright © 2020. Published by Elsevier Ltd.
Similar articles
-
Commissioning of beam shaper applicator for conformal intraoperative electron radiotherapy.Appl Radiat Isot. 2017 May;123:69-81. doi: 10.1016/j.apradiso.2017.02.039. Epub 2017 Feb 24. Appl Radiat Isot. 2017. PMID: 28260609
-
3D Monte Carlo dosimetry of intraoperative electron radiation therapy (IOERT).Phys Med. 2019 Jan;57:207-214. doi: 10.1016/j.ejmp.2018.12.037. Epub 2019 Jan 15. Phys Med. 2019. PMID: 30738527
-
IOeRT conventional and FLASH treatment planning system implementation exploiting fast GPU Monte Carlo: The case of breast cancer.Phys Med. 2024 May;121:103346. doi: 10.1016/j.ejmp.2024.103346. Epub 2024 Apr 11. Phys Med. 2024. PMID: 38608421
-
Evaluation of dosimetric properties of shielding disk used in intraoperative electron radiotherapy: A Monte Carlo study.Appl Radiat Isot. 2018 Sep;139:107-113. doi: 10.1016/j.apradiso.2018.04.037. Epub 2018 May 1. Appl Radiat Isot. 2018. PMID: 29751323
-
High-density dental implants and radiotherapy planning: evaluation of effects on dose distribution using pencil beam convolution algorithm and Monte Carlo method.J Appl Clin Med Phys. 2015 Sep 8;16(5):46–52. doi: 10.1120/jacmp.v16i5.5612. J Appl Clin Med Phys. 2015. PMID: 26699323 Free PMC article. Review.
Cited by
-
Delivery of intensity-modulated electron therapy by mechanical scanning: An algorithm study.Front Oncol. 2022 Nov 24;12:1063577. doi: 10.3389/fonc.2022.1063577. eCollection 2022. Front Oncol. 2022. PMID: 36505866 Free PMC article.
MeSH terms
LinkOut - more resources
Full Text Sources