Purpose: Antral pulmonary vein isolation (PVI) using a radiofrequency hot balloon catheter (RHB) is a feasible therapeutic option for treating atrial fibrillation (AF). Lesion durability after antral RHB-PVI remains unknown. This study aimed to evaluate lesion durability and the associations between procedural characteristics after antral RHB-PVI.
Methods: A total of 200 patients received antral RHB-PVI using the single-shot technique between April 2016 and March 2018. Antral RHB-PVI was performed following a pre-specified protocol and RHB energy application was performed for a maximum of two times for each PV. Consecutive patients who underwent repeated procedures for recurrence of any type of atrial tachyarrhythmia were enrolled.
Results: Twenty-six (13%) patients underwent repeated ablation and 20 patient documented AF recurrence (AF group) and 6 patients documented non-AF recurrence (non-AF group). Repeated ablation was performed at a median (25th, 75th percentiles) of 378 days (217, 487) after the initial procedure and durable PVI was observed in 86 (83.5%) PVs. Durability reached 89.7% when PVI was achieved only using an RHB. In the AF group (79 PVs), durable PVI was observed in 62 (78.5%) PVs, whereas all 24 PVs were still isolated in the non-AF group. The majority of reconnection sites were around the superior PVs.
Conclusions: Antral RHB-PVI shows high lesion durability, especially with both inferior PVs.
Keywords: Antral pulmonary vein isolation; Atrial fibrillation; Durability; Radiofrequency hot balloon ablation; Reconnection site.
© 2020. Springer Science+Business Media, LLC, part of Springer Nature.
Similar articles
-
Initial Result of Antrum Pulmonary Vein Isolation Using the Radiofrequency Hot-Balloon Catheter With Single-Shot Technique.JACC Clin Electrophysiol. 2019 Mar;5(3):354-363. doi: 10.1016/j.jacep.2019.01.017. JACC Clin Electrophysiol. 2019. PMID: 30898239
-
Validation of lesion durability following pulmonary vein isolation using the new third-generation laser balloon catheter in patients with recurrent atrial fibrillation.J Cardiol. 2021 Nov;78(5):388-396. doi: 10.1016/j.jjcc.2021.07.001. Epub 2021 Jul 29. J Cardiol. 2021. PMID: 34332839
-
Is the time of atrial fibrillation recurrence after cryoballoon ablation of paroxysmal atrial fibrillation influenced by the pattern of PV reconnections?J Interv Card Electrophysiol. 2021 Mar;60(2):321-327. doi: 10.1007/s10840-020-00804-z. Epub 2020 Jul 3. J Interv Card Electrophysiol. 2021. PMID: 32621212
-
Our approach to maximizing the durability of pulmonary vein isolation during a paroxysmal atrial fibrillation ablation procedure.J Cardiovasc Electrophysiol. 2012 Nov;23(11):1272-6. doi: 10.1111/j.1540-8167.2012.02414.x. Epub 2012 Aug 22. J Cardiovasc Electrophysiol. 2012. PMID: 22913515 Review.
-
Catheter Ablation of Paroxysmal Atrial Fibrillation: Have We Achieved Cure with Pulmonary Vein Isolation?Methodist Debakey Cardiovasc J. 2015 Apr-Jun;11(2):71-5. doi: 10.14797/mdcj-11-2-71. Methodist Debakey Cardiovasc J. 2015. PMID: 26306122 Free PMC article. Review.
Cited by
-
Current Balloon Devices for Ablation of Atrial Fibrillation.Rev Cardiovasc Med. 2024 Jan 22;25(1):34. doi: 10.31083/j.rcm2501034. eCollection 2024 Jan. Rev Cardiovasc Med. 2024. PMID: 39077656 Free PMC article. Review.
-
Effects of Hot Balloon vs. Cryoballoon Ablation for Atrial Fibrillation: A Systematic Review, Meta-Analysis, and Meta-Regression.Front Cardiovasc Med. 2021 Dec 15;8:787270. doi: 10.3389/fcvm.2021.787270. eCollection 2021. Front Cardiovasc Med. 2021. PMID: 34977192 Free PMC article.
References
-
- Reddy VY, Sediva L, Petru J, Skoda J, Chovanec M, Chitovova Z, et al. Durability of pulmonary vein isolation with cryoballoon ablation: results from the Sustained PV Isolation with Arctic Front Advance (SUPIR) study. J Cardiovasc Electrophysiol. 2015;26:493–500. https://doi.org/10.1111/jce.12626 . - DOI - PubMed
-
- Dukkipati SR, Neuzil P, Kautzner J, Petru J, Wichterle D, Skoda J, et al. The durability of pulmonary vein isolation using the visually guided laser balloon catheter: multicenter results of pulmonary vein remapping studies. Heart Rhythm. 2012;9:919–25. https://doi.org/10.1016/j.hrthm.2012.01.019 . - DOI - PubMed
-
- Sohara H, Ohe T, Okumura K, Naito S, Hirao K, Shoda M, et al. HotBalloon ablation of the pulmonary veins for paroxysmal AF: a multicenter randomized trial in Japan. J Am Coll Cardiol. 2016;68:2747–57. https://doi.org/10.1016/j.jacc.2016.10.037 . - DOI - PubMed
-
- Yamasaki H, Aonuma K, Shinoda Y, Komatsu Y, Masuda K, Hashimoto N, et al. Initial result of antrum pulmonary vein isolation using the radiofrequency hot-balloon catheter with single shot technique. JACC Clin Electrophysiol. 2019;5:354–63. https://doi.org/10.1016/j.jacep.2019.01.017 . - DOI - PubMed
-
- Kiuchi K, Kircher S, Watanabe N, Gaspar T, Rolf S, Arya A, et al. Quantitative analysis of isolation area and rhythm outcome in patients with paroxysmal atrial fibrillation after circumferential pulmonary vein antrum isolation using the pace-and-ablate technique. Circ Arrhythm Electrophysiol. 2012;5:667–75. https://doi.org/10.1161/CIRCEP.111.969923 . - DOI - PubMed
MeSH terms
LinkOut - more resources
Full Text Sources
Medical
Research Materials
Miscellaneous