Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
. 2021 Jun;35(3):319-323.
doi: 10.1080/02688697.2020.1812519. Epub 2020 Sep 17.

A comparison of the efficacy, safety, and duration of frame-based and Remebot robot-assisted frameless stereotactic biopsy

Affiliations

A comparison of the efficacy, safety, and duration of frame-based and Remebot robot-assisted frameless stereotactic biopsy

Shiqiang Wu et al. Br J Neurosurg. 2021 Jun.

Abstract

Objective: The aim of this study was to compare the efficacy, safety, and duration of Remebot robot-assisted frameless brain biopsy with those of standard frame-based stereotactic biopsy.

Patients and methods: A retrospective analysis of 66 patients undergoing stereotactic brain biopsy in our department from January 2015 to January 2019 was performed. We divided the patients into two groups: the frame-based group (n = 35) and the Remebot robot group (n = 31). Data on clinical characteristics, total procedure length, overall discomfort, diagnostic yield, complications, and postoperative length of hospital stay were retrospectively reviewed and compared between these two groups.

Results: No significant difference in diagnostic yield was detected in the two groups, with frame-based biopsy having a diagnostic yield of 91.4% and Remebot robot-assisted frameless brain biopsy having a diagnostic yield of 93.5%. The duration of the total procedure was 116.5 min for the frame-based biopsy and 80.1 min for the Remebot robot-assisted frameless brain biopsy (p < 0.001). There were no statistically significant differences in complication rate or postoperative duration of hospitalization between the two groups. The overall patient discomfort in the frame-based group was significantly greater than that in the Remebot robot group (visual analog scale score 2.7 ± 1.2 versus 1.5 ± 0.7, p = 0.001).

Conclusions: Remebot robot-assisted frameless brain biopsy was as efficacious and safe as standard stereotactic frame-based biopsy. However, frameless biopsy can alleviate the suffering of the patient and reduce the total duration of the procedure. Remebot robot-assisted frameless brain biopsy is easy to use and better accepted by patients than frame-based biopsy.

Keywords: Brain biopsy; Remebot robot; frame-based stereotaxy; frameless stereotaxy.

PubMed Disclaimer

Similar articles

Cited by

LinkOut - more resources