Preliminary Report on a Novel Virtual Reality Perimeter Compared With Standard Automated Perimetry
- PMID: 32941320
- DOI: 10.1097/IJG.0000000000001670
Preliminary Report on a Novel Virtual Reality Perimeter Compared With Standard Automated Perimetry
Abstract
Precis: The VisuALL head-mounted perimetry in normal subjects and glaucoma patients had a moderate to strong correlation with the Humphrey Field Analyzer (HFA).
Purpose: Visual field testing has a vital role in diagnosing and managing glaucoma. The current clinical practice relies on large, table-based testing units. This study investigated the performance of a novel virtual reality head-mounted visual perimetry device (VisuALL), in normal and glaucoma patients.
Methods: This prospective observational study was conducted on 50 eyes of 25 healthy subjects (normal group) and 52 eyes of 26 patients with a controlled mild or moderate stage of glaucoma (glaucoma group). All participants had visual field testing with VisuALL and the HFA (24-2, Swedish Interactive Threshold Algorithm). The mean sensitivity of the whole visual field and each quadrant were compared between both machines and the receiver operating characteristic was used to compare the diagnostic abilities and the Bland-Altman plot to evaluate the agreement of the 2 perimeters.
Results: The global mean sensitivity of the VisuALL and the HFA correlated significantly in both normal (r=0.5, P=0.001) and glaucoma (r=0.8, P<0.001) groups. The mean sensitivity of all quadrants also correlated significantly in both groups. The VisuALL mean sensitivity had a greater (0.98) receiver operating characteristic curve than HFA (0.93) mean sensitivity (P=0.06) in discriminating normal versus glaucoma.
Conclusion: There was an excellent correlation between the VisuALL and the Standard Automated Perimetry in normal and glaucoma patients and VisuALL showing high diagnostic performance.
Copyright © 2020 Wolters Kluwer Health, Inc. All rights reserved.
Conflict of interest statement
Disclosure: J.S.M. is a consultant at Aerie, Allergan, Glaukos, Olleyes; speaker at Aerie, Allergan; and received research support from Aerie, Allergan, Diopsys, Glaukos, Haag Streit, Nicox, Olleyes. L.J.K. received grant/research support from Allergan, Diopsys, Heidelberg Engineering, Alcon, Zeiss, Olleyes; speaker at Allergan, Glaukos, Bausch & Lomb and stock shareholder at Glaukos, Mati Therapeutics, Aerie, Olleyes. A.G.-G. declares no conflict of interest.
Comment in
-
Performance of virtual reality game-based automated perimetry in patients with childhood glaucoma.J AAPOS. 2023 Dec;27(6):325.e1-325.e6. doi: 10.1016/j.jaapos.2023.08.014. Epub 2023 Oct 30. J AAPOS. 2023. PMID: 38597674
References
-
- Wu Z, Medeiros FA. Recent developments in visual field testing for glaucoma. Curr Opin Ophthalmol. 2018;29:141–146.
-
- Glen FC, Baker H, Crabb DP. A qualitative investigation into patients’ views on visual field testing for glaucoma monitoring. BMJ Open. 2014;4:e003996.
-
- Werner EB, Petrig B, Krupin T, et al. Variability of automated visual fields in clinically stable glaucoma patients. Invest Ophthalmol Vis Sci. 1989;30:1083–1089.
-
- Lalle PA, Fingeret M, Eiden SB. Automated perimetry in the management of glaucoma. J Am Optom Assoc. 1989;60:900–911.
-
- Johnson CA, Wall M, Thompson HS, et al. A history of perimetry and visual field testing automated perimetry in the management of glaucoma. Optom Vis Sci. 2011;88:E8–E15.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical