PSG Validation of minute-to-minute scoring for sleep and wake periods in a consumer wearable device
- PMID: 32941498
- PMCID: PMC7498244
- DOI: 10.1371/journal.pone.0238464
PSG Validation of minute-to-minute scoring for sleep and wake periods in a consumer wearable device
Abstract
Background: Actigraphs are wrist-worn devices that record tri-axial accelerometry data used clinically and in research studies. The expense of research-grade actigraphs, however, limit their widespread adoption, especially in clinical settings. Tri-axial accelerometer-based consumer wearable devices have gained worldwide popularity and hold potential for a cost-effective alternative. The lack of independent validation of minute-to-minute accelerometer data with polysomnographic data or even research-grade actigraphs, as well as access to raw data has hindered the utility and acceptance of consumer-grade actigraphs.
Methods: Sleep clinic patients wore a consumer-grade wearable (Huami Arc) on their non-dominant wrist while undergoing an overnight polysomnography (PSG) study. The sample was split into two, 20 in a training group and 21 in a testing group. In addition to the Arc, the testing group also wore a research-grade actigraph (Philips Actiwatch Spectrum). Sleep was scored for each 60-s epoch on both devices using the Cole-Kripke algorithm.
Results: Based on analysis of our training group, Arc and PSG data were aligned best when a threshold of 10 units was used to examine the Arc data. Using this threshold value in our testing group, the Arc has an accuracy of 90.3%±4.3%, sleep sensitivity (or wake specificity) of 95.5%±3.5%, and sleep specificity (wake sensitivity) of 55.6%±22.7%. Compared to PSG, Actiwatch has an accuracy of 88.7%±4.5%, sleep sensitivity of 92.6%±5.2%, and sleep specificity of 60.5%±20.2%, comparable to that observed in the Arc.
Conclusions: An optimized sleep/wake threshold value was identified for a consumer-grade wearable Arc trained by PSG data. By applying this sleep/wake threshold value for Arc generated accelerometer data, when compared to PSG, sleep and wake estimates were adequate and comparable to those generated by a clinical-grade actigraph. As with other actigraphs, sleep specificity plateaus due to limitations in distinguishing wake without movement from sleep. Further studies are needed to evaluate the Arc's ability to differentiate between sleep and wake using other sources of data available from the Arc, such as high resolution accelerometry and photoplethysmography.
Conflict of interest statement
The authors have declared that no competing interests exist.
Figures




Similar articles
-
Validation of a Consumer Sleep Wearable Device With Actigraphy and Polysomnography in Adolescents Across Sleep Opportunity Manipulations.J Clin Sleep Med. 2019 Sep 15;15(9):1337-1346. doi: 10.5664/jcsm.7932. J Clin Sleep Med. 2019. PMID: 31538605 Free PMC article.
-
Developing a Sleep Algxorithm to Support a Digital Medicine System: Noninterventional, Observational Sleep Study.JMIR Ment Health. 2024 Dec 20;11:e62959. doi: 10.2196/62959. JMIR Ment Health. 2024. PMID: 39727095 Free PMC article.
-
Validation of the Sleep-Wake Scoring of a New Wrist-Worn Sleep Monitoring Device.J Clin Sleep Med. 2018 Jun 15;14(6):1057-1062. doi: 10.5664/jcsm.7180. J Clin Sleep Med. 2018. PMID: 29852899 Free PMC article.
-
A systematic review of the accuracy of sleep wearable devices for estimating sleep onset.Sleep Med Rev. 2020 Feb;49:101227. doi: 10.1016/j.smrv.2019.101227. Epub 2019 Nov 6. Sleep Med Rev. 2020. PMID: 31901524
-
Consumer sleep tracking devices: a review of mechanisms, validity and utility.Expert Rev Med Devices. 2016 May;13(5):497-506. doi: 10.1586/17434440.2016.1171708. Epub 2016 Apr 18. Expert Rev Med Devices. 2016. PMID: 27043070 Review.
Cited by
-
Validity of Two Consumer Multisport Activity Tracker and One Accelerometer against Polysomnography for Measuring Sleep Parameters and Vital Data in a Laboratory Setting in Sleep Patients.Sensors (Basel). 2022 Dec 6;22(23):9540. doi: 10.3390/s22239540. Sensors (Basel). 2022. PMID: 36502241 Free PMC article.
-
Performance of consumer wrist-worn sleep tracking devices compared to polysomnography: a meta-analysis.J Clin Sleep Med. 2025 Mar 1;21(3):573-582. doi: 10.5664/jcsm.11460. J Clin Sleep Med. 2025. PMID: 39484805
-
Relationship Analogy between Sleep Bruxism and Temporomandibular Disorders in Children: A Narrative Review.Children (Basel). 2022 Sep 25;9(10):1466. doi: 10.3390/children9101466. Children (Basel). 2022. PMID: 36291402 Free PMC article. Review.
-
Association Between Nocturnal Sleep Duration and Obesity Indicators Among People with Type 2 Diabetes: A Cross-Sectional Study in Ningbo, China.Diabetes Metab Syndr Obes. 2022 May 3;15:1357-1364. doi: 10.2147/DMSO.S350347. eCollection 2022. Diabetes Metab Syndr Obes. 2022. PMID: 35535215 Free PMC article.
-
Intermittent Fasting and Sleep: A Review of Human Trials.Nutrients. 2021 Oct 1;13(10):3489. doi: 10.3390/nu13103489. Nutrients. 2021. PMID: 34684490 Free PMC article. Review.
References
Publication types
MeSH terms
Associated data
Grants and funding
LinkOut - more resources
Full Text Sources
Research Materials