Utilization of Intravascular Ultrasound and Aptus Heli-Fx EndoAnchors for a Complex Type Ia Endoleak Repair
- PMID: 32947001
- DOI: 10.1016/j.avsg.2020.09.012
Utilization of Intravascular Ultrasound and Aptus Heli-Fx EndoAnchors for a Complex Type Ia Endoleak Repair
Abstract
Type Ia endoleaks are a complication of endovascular aneurysm repair that require intervention due to the increased risk of aneurysmal rupture. Many techniques have been studied in order to minimize the occurrence of these complications, as well as to aid in their repair. We introduce the utilization of intravascular ultrasound in conjunction with the endovascular EndoAnchor systems for the repair of a complex type Ia endoleak in an infrarenal aortic stent graft. This technique allowed us to have real-time feedback of EndoAnchor placement and aortic wall to stent graft apposition with a reduction in radiation exposure and contrast dye load.
Copyright © 2020 Elsevier Inc. All rights reserved.