Yield, clinical significance, and cost of a combination BACTEC plus Septi-Chek blood culture system
- PMID: 3294885
- PMCID: PMC266095
- DOI: 10.1128/jcm.25.5.819-823.1987
Yield, clinical significance, and cost of a combination BACTEC plus Septi-Chek blood culture system
Abstract
A blood culture was performed by adding a vented Septi-Chek bottle (Roche Diagnostics, Div. Hoffmann-LaRoche Inc., Nutley, N.J.) to a standard BACTEC system (Johnston Laboratories, Inc., Towson, Md.) blood culture. The yield of bacteremic patients, the clinical significance of organisms detected, and the cost of the combination system were compared with those of the standard BACTEC system alone. Each culture included 20 ml of blood divided among a BACTEC 6B aerobic bottle (5 ml), a BACTEC 7D anaerobic bottle (5 ml), and a Septi-Chek bottle equipped with a slide subculture attachment (10 ml). Significant isolates grew in 9.6% of the 2,269 cultures evaluated. The combination BACTEC plus Septi-Chek system detected 25% more bacteremic patients than the BACTEC system alone, 129 patients versus 103. The 26 bacteremic patients detected by only the added Septi-Chek bottle included 7 whose organism was isolated from blood alone and 19 whose organism was in mixed or pure culture from a second source. Detection of the organism resulted in alteration of antimicrobial therapy in 17 of these 26 patients. The combination system, which cultured a 20-ml blood volume, cost $11,000 more during the study period than the BACTEC system alone, which cultured a 10-ml volume. Reimbursement under the diagnosis-related group system was increased by $23,000 as a result of documentation of sepsis in these 26 patients. Blood volume and, possibly, the use of multiple blood culture systems are important factors when selecting a blood culture procedure for routine use.
Similar articles
-
Comparative evaluation of BACTEC aerobic Plus/F and Septi-Chek Release blood culture media.J Clin Microbiol. 1996 Jan;34(1):126-9. doi: 10.1128/jcm.34.1.126-129.1996. J Clin Microbiol. 1996. PMID: 8748287 Free PMC article.
-
Controlled evaluation of BACTEC Plus 26 and Roche Septi-Chek aerobic blood culture bottles.J Clin Microbiol. 1991 May;29(5):879-82. doi: 10.1128/jcm.29.5.879-882.1991. J Clin Microbiol. 1991. PMID: 2056053 Free PMC article.
-
Controlled evaluation of the agar-slide and radiometric blood culture systems for the detection of bacteremia and fungemia.J Clin Microbiol. 1986 Feb;23(2):221-5. doi: 10.1128/jcm.23.2.221-225.1986. J Clin Microbiol. 1986. PMID: 3517047 Free PMC article.
-
Controlled comparison of a new Becton Dickinson agar slant blood culture system with Roche Septi-Chek for the detection of bacteremia and fungemia.J Clin Microbiol. 1989 Dec;27(12):2637-9. doi: 10.1128/jcm.27.12.2637-2639.1989. J Clin Microbiol. 1989. PMID: 2687314 Free PMC article.
-
[Laboratory media for the cultivation of tubercle bacillus].Kekkaku. 1998 May;73(5):329-37. Kekkaku. 1998. PMID: 9637817 Review. Japanese.
Cited by
-
Evolving approaches to management of quality in clinical microbiology.Clin Microbiol Rev. 1994 Jan;7(1):55-88. doi: 10.1128/CMR.7.1.55. Clin Microbiol Rev. 1994. PMID: 8118791 Free PMC article. Review.
References
MeSH terms
LinkOut - more resources
Full Text Sources
Medical