. 2020 Nov;32(11):1470-1474.
doi: 10.1097/MEG.0000000000001929.

Endoscopic ultrasound-guided liver biopsy using a 20-gauge fine needle biopsy needle with the wet-heparinized suction technique

Affiliations

Endoscopic ultrasound-guided liver biopsy using a 20-gauge fine needle biopsy needle with the wet-heparinized suction technique

Gurhan Sisman et al. Eur J Gastroenterol Hepatol. 2020 Nov.

Abstract

Objectives: The primary aim of this study was to evaluate the histological adequacy of the liver tissue specimens obtained with a 20-gauge fine-needle biopsy needle and the secondary aim was to test the safety endoscopic ultrasound-guided liver biopsy with a 20-gauge fine-needle biopsy needle with the wet-heparinized suction technique.

Methods: Forty patients who underwent endoscopic ultrasound-guided liver biopsy were included in the study. A 20-gauge fine-needle biopsy needle was used with the wet-heparinized suction technique to make one pass each from the left and the right lobe. Histologic characteristics of the specimens were evaluated, and patients were observed after the procedure in order to intervene in case of an adverse event.

Results: The median longest core fragment was 22 mm from the left lobe [first quartile-third quartile 20-25 mm, interquartile range (IQR) 5 mm], and 20 mm (first quartile-third quartile 17-22 mm, IQR 5 mm) from the right lobe. The median cumulative core length per patient was 103 mm (91-108 mm, IQR 17 mm). The median cumulative number of complete portal triads per patient was 69.50 (52.25-82.25, IQR 30). The rate of diagnostic yield was 100%. Post-biopsy self-limiting abdominal pain was reported in two patients (5%). The most common histologic diagnosis was fatty liver disease (25%).

Conclusion: Endoscopic ultrasound-guided liver biopsy with the wet-heparinized suction technique using a 20-gauge fine-needle biopsy needle is a safe alternative method in clinical practice.

PubMed Disclaimer

References

    1. Younossi ZM, Koenig AB, Abdelatif D, Fazel Y, Henry L, Wymer M. Global epidemiology of nonalcoholic fatty liver disease-meta-analytic assessment of prevalence, incidence, and outcomes. Hepatology. 2016; 64:73–84
    1. Skelly MM, James PD, Ryder SD. Findings on liver biopsy to investigate abnormal liver function tests in the absence of diagnostic serology. J Hepatol. 2001; 35:195–199
    1. Mohan BP, Shakhatreh M, Garg R, Ponnada S, Adler DG. Efficacy and safety of EUS-guided liver biopsy: a systematic review and meta-analysis. Gastrointest Endosc. 2019; 89:238–246.e3
    1. Mok SRS, Diehl DL. The role of EUS in liver biopsy. Curr Gastroenterol Rep. 2019; 21:6
    1. Mok SRS, Diehl DL, Johal AS, Khara HS, Confer BD, Mudireddy PR, et al. A prospective pilot comparison of wet and dry heparinized suction for EUS-guided liver biopsy (with videos). Gastrointest Endosc. 2018; 88:919–925

LinkOut - more resources