. 2021 Apr;31(4):2106-2114.
doi: 10.1007/s00330-020-07290-x. Epub 2020 Sep 21.

Radiation dose management systems-requirements and recommendations for users from the ESR EuroSafe Imaging initiative

Affiliations

Radiation dose management systems-requirements and recommendations for users from the ESR EuroSafe Imaging initiative

Reinhard W Loose et al. Eur Radiol. 2021 Apr.

Abstract

The European Directive 2013/59/Euratom requires member states of the European Union to ensure justification and optimisation of radiological procedures and store information on patient exposure for analysis and quality assurance. The EuroSafe Imaging campaign of the European Society of Radiology created a working group (WG) on "Dose Management" with the aim to provide European recommendations on the implementation of dose management systems (DMS) in clinical practice. The WG follows Action 4: "Promote dose management systems to establish local, national, and European diagnostic reference levels (DRL)" of the EuroSafe Imaging Call for Action 2018. DMS are designed for medical practitioners, radiographers, medical physics experts (MPE) and other health professionals involved in imaging to support their tasks and duties of radiation protection in accordance with local and national requirements. The WG analysed requirements and critical points when installing a DMS and classified the individual functions at different performance levels. KEY POINTS: • DMS are very helpful software tools for monitoring patient exposure, optimisation, compliance with DRLs and quality assurance. • DMS can help to fulfil dosimetric aspects of the European Directive 2013/59/Euratom. • The EuroSafe WG analyses DMS requirements and gives recommendations for users.

Keywords: Dose management systems; Optimisation; Quality assurance; Radiation protection.

PubMed Disclaimer

Conflict of interest statement

The authors of this manuscript declare no relationships with any companies whose products or services may be related to the subject matter of the article.

Figures

Fig. 1
Fig. 1
Workflow of dose management (DMS) in an environment of HIS, RIS, PACS and modalities (B. Renger, Munich, 2019)

References

    1. European Council Directive 2013/59/Euratom on basic safety standards for protection against the dangers arising from exposure to ionising radiation (2014). Official Journal of the European Union L13;57:1-73 Available at https://eur-lex.europa.eu/LexUriServ/LexUriServ.do?uri=OJ:L:2014:013:000.... Accessed 13 Jan 2020
    1. European Society of Radiology (ESR) Summary of the European Directive 2013/59/Euratom: essentials for health professionals in radiology. Insights Imaging. 2015;6:411–417. doi: 10.1007/s13244-015-0410-4. - DOI - PMC - PubMed
    1. EuroSafe Imaging Call for Action 2018, http://www.eurosafeimaging.org/about/call-for-action Accessed 13 Jan 2020
    1. Vañó E, Miller DL, Martin CJ et al (2017) ICRP Publication 135: Diagnostic reference levels in medical imaging. Ann ICRP 46(1):1–144. 10.1177/0146645317717209 - PubMed
    1. Radiation Protection No. 185 (2018) European guidelines on diagnostic reference levels for paediatric imaging. Eur Commission. 10.2833/486256