Cuff Under Pressure for Greater Accuracy
- PMID: 32959103
- DOI: 10.1007/s11906-020-01103-8
Cuff Under Pressure for Greater Accuracy
Abstract
Purpose of review: To present the evidence that describes what is being measured by upper-arm cuff blood pressure (BP) and the level of accuracy compared with invasive central aortic and brachial BP. Potential causes of inaccuracy and emerging methods are also discussed.
Recent findings: On average cuff systolic BP systematically underestimates invasive brachial systolic BP, although in a given individual it may substantially under- or over-estimate central aortic systolic BP. Such errors may affect individual health management outcomes and distort population level data on hypertension prevalence and control. Oscillometric cuff BP is particularly susceptible to inaccuracy in people with high arterial stiffness and with pathophysiological BP waveform shapes. Emerging cuff-less BP methods will be susceptible to inaccuracy if oscillometric cuff BP is used for calibration. The original purpose of cuff BP was to estimate central aortic BP. Recent evidence has shown substantial inaccuracy of oscillometric cuff BP exists for the measurement of invasive central aortic and brachial BP. Thus, development of more accurate BP methods, through better understanding of oscillometric and BP waveform morphology, is needed to improve health outcomes related to high BP.
Keywords: Aortic blood pressure; BP; BP management; Blood pressure cuff; Brachial blood pressure; Hypertension.
Similar articles
-
Validation Study to Determine the Accuracy of Central Blood Pressure Measurement Using the Sphygmocor Xcel Cuff Device.Hypertension. 2020 Jul;76(1):244-250. doi: 10.1161/HYPERTENSIONAHA.120.14916. Epub 2020 Jun 1. Hypertension. 2020. PMID: 32475318
-
Estimating central systolic blood pressure during oscillometric determination of blood pressure: proof of concept and validation by comparison with intra-aortic pressure recording and arterial tonometry.Blood Press Monit. 2012 Jun;17(3):132-6. doi: 10.1097/MBP.0b013e328352ae5b. Blood Press Monit. 2012. PMID: 22466804
-
Central Blood Pressure Monitoring via a Standard Automatic Arm Cuff.Sci Rep. 2017 Oct 31;7(1):14441. doi: 10.1038/s41598-017-14844-5. Sci Rep. 2017. PMID: 29089581 Free PMC article.
-
Accuracy of Cuff-Measured Blood Pressure: Systematic Reviews and Meta-Analyses.J Am Coll Cardiol. 2017 Aug 1;70(5):572-586. doi: 10.1016/j.jacc.2017.05.064. J Am Coll Cardiol. 2017. PMID: 28750701 Review.
-
Arterial pressure waveforms in hypertension.Minerva Med. 2003 Aug;94(4):229-50. Minerva Med. 2003. PMID: 14605588 Review.
Cited by
-
Direct estimation of central aortic pressure from measured or quantified mean and diastolic brachial blood pressure: agreement with invasive records.Front Cardiovasc Med. 2023 Jul 25;10:1207069. doi: 10.3389/fcvm.2023.1207069. eCollection 2023. Front Cardiovasc Med. 2023. PMID: 37560119 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical
Research Materials