Comparison of two-dimensional high-definition, ultra high-definition and three-dimensional endovision systems: an ex-vivo randomised study
- PMID: 32959182
- DOI: 10.1007/s00464-020-07980-z
Comparison of two-dimensional high-definition, ultra high-definition and three-dimensional endovision systems: an ex-vivo randomised study
Abstract
Background: Two-dimensional high-definition (2D HD) endovision system is preferred for laparoscopic surgery. Recently, new generation three-dimensional (3D) HD and ultra-HD (4K) endovision systems are introduced to improve the safety and efficacy of laparoscopic surgery. There is limited evidence on superiority of one technology over the others. This experimental trial was designed to evaluate 2D HD, 3D HD and 4K HD endovision systems in performance of standardized tasks.
Methods: This was a randomized, cross-over experimental study. Twenty-one surgical residents who were exposed to laparoscopic surgery were enrolled. Participants were randomly assigned into three groups. Each group performed standardised tasks i.e. peg transfer, precision cutting, navigating in space and intra-corporeal suturing using 2D HD, 4K HD and 3D HD endovision systems on a box trainer. Procedures were recorded as 2D HD videos and analysed later. Participant's perceived workload was assessed using Surg-TLX questionnaire. Primary endpoints were execution time in seconds and error score. Secondary endpoint was workload assessment.
Results: The 3D HD had shorter execution time compared to 2D HD and 4K HD in all tasks except precision cutting (p = 0.004, 0.03, 0.001, 0.001 and p = 0.002, 0.191, 0.006, 0.005 in peg transfer, precision cutting, navigating in space and intra-corporeal suturing respectively). The 4K HD was significantly faster than 2D HD only in navigating in space task (p = 0.002). The error score between 3D HD and 4K HD were comparable in all tasks. The 2D HD had significantly more error scores compared to 4K HD, 3D HD in peg transfer task (p = 0.005, 0.014, respectively). 3D HD had significantly less workload than 2D HD and 4K HD in most of the dimensions of Surg-TLX CONCLUSIONS: 3D HD endovision system in comparison to 2D HD and 4K HD, may lead to faster execution without compromising safety of a task and is associated with less workload.
Keywords: 2D HD; 3D HD; 4K HD; Endovision system; Laparoscopy; Surg-TLX.
© 2020. Springer Science+Business Media, LLC, part of Springer Nature.
References
-
- Pierre SA, Ferrandino MN, Simmons WN, Fernandez C, Zhong P, Albala DM, Preminger GM (2009) High definition laparoscopy: objective assessment of performance characteristics and comparison with standard laparoscopy. J Endourol 23:523–528 - DOI
-
- Ukai K, Howarth PA (2008) Visual fatigue caused by viewing stereoscopic motion images: Background, theories, and observations. Displays 29:106–116 - DOI
-
- Storz P, Buess GF, Kunert W, Kirschniak A (2012) 3D HD versus 2D HD: surgical task efficiency in standardised phantom tasks. Surg Endosc 26:1454–1460 - DOI
-
- Wagner M, Mayer BFB, Bodenstedt S, Stemmer K, Fereydooni A, Speidel S, Dillmann R, Nickel F, Fischer L, Kenngott HG (2018) Computer-assisted 3D bowel length measurement for quantitative laparoscopy. Surg Endosc 32:4052–4061 - DOI
-
- Sakata S, Grove PM, Hill A, Watson MO, Stevenson ARL (2017) Impact of simulated three-dimensional perception on precision of depth judgements, technical performance and perceived workload in laparoscopy. Br J Surg 104:1097–1106 - DOI
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources