. 2021 Feb 1;26(1):65-69.
doi: 10.1097/MBP.0000000000000484.

Performance of an automated blood pressure measurement device in a stroke rehabilitation unit

Affiliations

Performance of an automated blood pressure measurement device in a stroke rehabilitation unit

Maxime Lamarre-Cliche et al. Blood Press Monit. .

Abstract

Context: Valid blood pressure (BP) measurements are needed in post-stroke rehabilitation hospital units for the management of hypertension. Automated devices could be used to improve on usual care BP measurement. However, more information is needed about the performance of these devices in such a context.

Methods: This prospective nonrandomized study was performed in stroke patients with hypertension hospitalized in a stroke rehabilitation unit. Two in-hospital BP assessment strategies were compared: usual care BP and in-hospital automated office BP (AOBP) standardized measurements. In-office AOBP and ambulatory BP monitoring (ABPM) were also performed on these patients. The main outcome was SBP. Study follow-up was until discharge, up to a maximum of 4 weeks.

Results: Sixty-two patients with stroke hospitalized in a rehabilitation unit were included. Usual care BP was 130 ± 12/79 ± 9 mmHg and differed from an in-hospital AOBP of 117 ± 14/75 ± 12 mmHg (P < 0.001/P < 0.001). In-hospital and in-office AOBP measurements did not differ. Twenty percent of patients reached SBP therapeutic goals according to in-hospital AOBP but not according to usual care BP measurements.

Conclusion: This study shows that in a post-stroke rehabilitation unit, standardized in-hospital AOBP estimates are on average much lower than the usual care BP correlates and similar to the in-office AOBP estimates. In-hospital AOBP devices in a stroke rehabilitation unit could add important information for the management of hypertension.

PubMed Disclaimer

References

    1. O’Donnell MJ, Xavier D, Liu L, Zhang H, Chin SL, Rao-Melacini P, et al.; INTERSTROKE investigators. Risk factors for ischaemic and intracerebral haemorrhagic stroke in 22 countries (the INTERSTROKE study): a case-control study. Lancet. 2010; 376:112–123.
    1. Ettehad D, Emdin CA, Kiran A, Anderson SG, Callender T, Emberson J, et al. Blood pressure lowering for prevention of cardiovascular disease and death: a systematic review and meta-analysis. Lancet. 2016; 387:957–967.
    1. PROGRESS Collaborative Group. Randomised trial of a perindopril-based blood-pressure-lowering regimen among 6,105 individuals with previous stroke or transient ischaemic attack. Lancet. 2001; 358:1033–1041.
    1. Turnbull F; Blood Pressure Lowering Treatment Trialists’ Collaboration. Effects of different blood-pressure-lowering regimens on major cardiovascular events: results of prospectively-designed overviews of randomised trials. Lancet. 2003; 362:1527–1535.
    1. Nerenberg KA, Zarnke KB, Leung AA, Dasgupta K, Butalia S, McBrien K, et al.; Hypertension Canada. Hypertension Canada’s 2018 guidelines for diagnosis, risk assessment, prevention, and treatment of hypertension in adults and children. Can J Cardiol. 2018; 34:506–525.