High-resolution multi-T1 -weighted contrast and T1 mapping with low sensitivity using the fluid and white matter suppression (FLAWS) sequence at 7T
- PMID: 32989788
- DOI: 10.1002/mrm.28517
High-resolution multi-T1 -weighted contrast and T1 mapping with low sensitivity using the fluid and white matter suppression (FLAWS) sequence at 7T
Abstract
Purpose: To demonstrate that fluid and white matter suppression (FLAWS) imaging can be used for high-resolution T1 mapping with low transmitted bias field ( ) sensitivity at 7T.
Methods: The FLAWS sequence was optimized for 0.8-mm isotropic resolution imaging. The theoretical accuracy and precision of the FLAWS T1 mapping was compared with the one of the magnetization-prepared two rapid gradient echoes (MP2RAGE) sequence optimized for low sensitivity. FLAWS images were acquired at 7T on six healthy volunteers (21 to 48 years old; two women). MP2RAGE and saturation-prepared with two rapid gradient echoes (SA2RAGE) datasets were also acquired to obtain T1 mapping references and maps. The contrast-to-noise ratio (CNR) between brain tissues was measured in the FLAWS-hco and MP2RAGE-uni images. The Pearson correlation was measured between the MP2RAGE and FLAWS T1 maps. The effect of on FLAWS T1 mapping was assessed using the Pearson correlation.
Results: The FLAWS-hco images were characterized by a higher brain tissue CNR ( , , ) than the MP2RAGE-uni images ( , , ). The theoretical accuracy and precision of the FLAWS T1 mapping ( ) were in agreement with those provided by the MP2RAGE T1 mapping ( ). A good agreement was found between in vivo T1 values measured with the MP2RAGE and FLAWS sequences (r = 0.91). A weak correlation was found between the FLAWS T1 map and the map within cortical gray matter and white matter segmentations ( ; ).
Conclusion: The results from this study suggest that FLAWS is a good candidate for high-resolution T1 -weighted imaging and T1 mapping at the field strength of 7T.
Keywords: Brain; FLAWS; MP2RAGE; MRI; T1 mapping.
© 2020 International Society for Magnetic Resonance in Medicine.
Similar articles
-
Advantages of fluid and white matter suppression (FLAWS) with MP2RAGE compared with double inversion recovery turbo spin echo (DIR-TSE) at 7T.Eur J Radiol. 2019 Jul;116:160-164. doi: 10.1016/j.ejrad.2019.04.019. Epub 2019 Apr 29. Eur J Radiol. 2019. PMID: 31153559
-
Fluid and White Matter Suppression: New Sensitive 3 T Magnetic Resonance Imaging Contrasts for Cortical Lesion Detection in Multiple Sclerosis.Invest Radiol. 2022 Sep 1;57(9):592-600. doi: 10.1097/RLI.0000000000000877. Epub 2022 May 3. Invest Radiol. 2022. PMID: 35510874 Free PMC article.
-
Optimization and validation of low-field MP2RAGE T1 mapping on 0.35T MR-Linac: Toward adaptive dose painting with hypoxia biomarkers.Med Phys. 2024 Nov;51(11):8124-8140. doi: 10.1002/mp.17353. Epub 2024 Aug 14. Med Phys. 2024. PMID: 39140821
-
Whole brain and deep gray matter structure segmentation: Quantitative comparison between MPRAGE and MP2RAGE sequences.PLoS One. 2021 Aug 6;16(8):e0254597. doi: 10.1371/journal.pone.0254597. eCollection 2021. PLoS One. 2021. PMID: 34358242 Free PMC article.
-
Malformations of cortical development: The role of 7-Tesla magnetic resonance imaging in diagnosis.Rev Neurol (Paris). 2019 Mar;175(3):157-162. doi: 10.1016/j.neurol.2019.01.393. Epub 2019 Mar 1. Rev Neurol (Paris). 2019. PMID: 30827579 Review.
Cited by
-
Improving the understanding and performance of clinical MRI using tissue property filters and the central contrast theorem, MASDIR pulse sequences and synergistic contrast MRI.Quant Imaging Med Surg. 2022 Sep;12(9):4658-4690. doi: 10.21037/qims-22-394. Quant Imaging Med Surg. 2022. PMID: 36060593 Free PMC article. Review.
-
Simultaneous Optimization of MP2RAGE T1 -weighted (UNI) and FLuid And White matter Suppression (FLAWS) brain images at 7T using Extended Phase Graph (EPG) Simulations.Magn Reson Med. 2023 Mar;89(3):937-950. doi: 10.1002/mrm.29479. Epub 2022 Nov 9. Magn Reson Med. 2023. PMID: 36352772 Free PMC article.
-
Ultra-High Contrast (UHC) MRI of the Brain, Spinal Cord and Optic Nerves in Multiple Sclerosis Using Directly Acquired and Synthetic Bipolar Filter (BLAIR) Images.Diagnostics (Basel). 2025 Jan 30;15(3):329. doi: 10.3390/diagnostics15030329. Diagnostics (Basel). 2025. PMID: 39941259 Free PMC article. Review.
-
Multi T1-weighted contrast imaging and T1 mapping with compressed sensing FLAWS at 3 T.MAGMA. 2023 Oct;36(5):823-836. doi: 10.1007/s10334-023-01071-5. Epub 2023 Feb 27. MAGMA. 2023. PMID: 36847989
-
Targeted magnetic resonance imaging (tMRI) of small changes in the T1 and spatial properties of normal or near normal appearing white and gray matter in disease of the brain using divided subtracted inversion recovery (dSIR) and divided reverse subtracted inversion recovery (drSIR) sequences.Quant Imaging Med Surg. 2023 Oct 1;13(10):7304-7337. doi: 10.21037/qims-23-232. Epub 2023 Aug 15. Quant Imaging Med Surg. 2023. PMID: 37869282 Free PMC article. Review.
References
REFERENCES
-
- Trattnig S, Springer E, Bogner W, et al. Key clinical benefits of neuroimaging at 7 T. Neuroimage. 2018;168:477-489.
-
- Marques JP, Kober T, Krueger G, van der Zwaag W, Van de Moortele P-F, Gruetter R. MP2RAGE, a self bias-field corrected sequence for improved segmentation and T1-mapping at high field. Neuroimage. 2010;49:1271-1281.
-
- Okubo G, Okada T, Yamamoto A, et al. MP2RAGE for deep gray matter measurement of the brain: A comparative study with MPRAGE. J Magn Reson Imaging. 2016;43:55-62.
-