High-resolution multi-T1 -weighted contrast and T1 mapping with low sensitivity using the fluid and white matter suppression (FLAWS) sequence at 7T
- PMID: 32989788
- DOI: 10.1002/mrm.28517
High-resolution multi-T1 -weighted contrast and T1 mapping with low sensitivity using the fluid and white matter suppression (FLAWS) sequence at 7T
Abstract
Purpose: To demonstrate that fluid and white matter suppression (FLAWS) imaging can be used for high-resolution T1 mapping with low transmitted bias field ( ) sensitivity at 7T.
Methods: The FLAWS sequence was optimized for 0.8-mm isotropic resolution imaging. The theoretical accuracy and precision of the FLAWS T1 mapping was compared with the one of the magnetization-prepared two rapid gradient echoes (MP2RAGE) sequence optimized for low sensitivity. FLAWS images were acquired at 7T on six healthy volunteers (21 to 48 years old; two women). MP2RAGE and saturation-prepared with two rapid gradient echoes (SA2RAGE) datasets were also acquired to obtain T1 mapping references and maps. The contrast-to-noise ratio (CNR) between brain tissues was measured in the FLAWS-hco and MP2RAGE-uni images. The Pearson correlation was measured between the MP2RAGE and FLAWS T1 maps. The effect of on FLAWS T1 mapping was assessed using the Pearson correlation.
Results: The FLAWS-hco images were characterized by a higher brain tissue CNR ( , , ) than the MP2RAGE-uni images ( , , ). The theoretical accuracy and precision of the FLAWS T1 mapping ( ) were in agreement with those provided by the MP2RAGE T1 mapping ( ). A good agreement was found between in vivo T1 values measured with the MP2RAGE and FLAWS sequences (r = 0.91). A weak correlation was found between the FLAWS T1 map and the map within cortical gray matter and white matter segmentations ( ; ).
Conclusion: The results from this study suggest that FLAWS is a good candidate for high-resolution T1 -weighted imaging and T1 mapping at the field strength of 7T.
Keywords: Brain; FLAWS; MP2RAGE; MRI; T1 mapping.
© 2020 International Society for Magnetic Resonance in Medicine.
References
REFERENCES
-
- Trattnig S, Springer E, Bogner W, et al. Key clinical benefits of neuroimaging at 7 T. Neuroimage. 2018;168:477-489.
-
- Marques JP, Kober T, Krueger G, van der Zwaag W, Van de Moortele P-F, Gruetter R. MP2RAGE, a self bias-field corrected sequence for improved segmentation and T1-mapping at high field. Neuroimage. 2010;49:1271-1281.
-
- Okubo G, Okada T, Yamamoto A, et al. MP2RAGE for deep gray matter measurement of the brain: A comparative study with MPRAGE. J Magn Reson Imaging. 2016;43:55-62.
-
- Kober T, Granziera C, Ribes D, et al. MP2RAGE multiple sclerosis magnetic resonance imaging at 3 T. Invest Radiol. 2012;47:346-352.
-
- Simioni S, Amarù F, Bonnier G, et al. MP2RAGE provides new clinically-compatible correlates of mild cognitive deficits in relapsing-remitting multiple sclerosis. J Neurol. 2014;261:1606-1613.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources