. 2021 Mar;22(2):80-84.
doi: 10.1080/14670100.2020.1824640. Epub 2020 Sep 30.

Skin flap thickness and magnet strength in Cochlear implants

Affiliations

Skin flap thickness and magnet strength in Cochlear implants

Tamara Searle et al. Cochlear Implants Int. 2021 Mar.

Abstract

Objectives: The new CI600 series cochlear implants (Nucleus® Profile™ Plus Series; CI) are more MRI (magnetic resonance imaging) compatible. The magnet's attraction force is lower, possibly posing a problem in patients needing a higher strength magnet due to increased skin thickness. This study aims to investigate the relationship between skin thickness and magnet strength in patients undergoing cochlear implantation. Methods: This cross-sectional study used data from 46 patients with cochlear implants. Data collected included age, body mass index, magnet strength used, pre-operative and post-operative imaging and skin thickness. Results: There was a weak, positive correlation between magnet strength and skin thickness, which was not statistically significant (r(46) = 0.149, p = 0.324). There was a weak, positive correlation between magnet strength and body mass index, which was not statistically significant (r(46) = 0.113, p = 0.456). There was a moderate, positive correlation between body mass index and skin thickness, which was statistically significant (r(46) = 0.362, p = 0.012). Discussion: The increased skin thickness in patients with higher body mass indexes, might mean that these patients will require a stronger magnet. Conclusion: There are concerns that the lower attraction force in the CI600 (Nucleus® Profile™ Plus Series) may pose a retention problem for these patients. This might mean that discussions about skin flap reduction surgery are needed.

Keywords: Body mass index; Cochlear implant; Computer tomography; Magnet retention; Magnet strength; Magnetic resonance imaging.