Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
. 2020 Dec:197:105758.
doi: 10.1016/j.cmpb.2020.105758. Epub 2020 Sep 15.

An introduction to the Cyrcadia Breast Monitor: A wearable breast health monitoring device

Affiliations

An introduction to the Cyrcadia Breast Monitor: A wearable breast health monitoring device

Vinitha Sree S et al. Comput Methods Programs Biomed. 2020 Dec.

Erratum in

Abstract

Background: The most common breast cancer detection modalities are generally limited by radiation exposure, discomfort, high costs, inter-observer variabilities in image interpretation, and low sensitivity in detecting cancer in dense breast tissue. Therefore, there is a clear need for an affordable and effective adjunct modality that can address these limitations. The Cyrcadia Breast Monitor (CBM) is a non-invasive, non-compressive, and non-radiogenic wearable device developed as an adjunct to current modalities to assist in the detection of breast tissue abnormalities in any type of breast tissue.

Methods: The CBM records thermodynamic metabolic data from the breast skin surface over a period of time using two wearable biometric patches consisting of eight sensors each and a data recording device. The acquired multi-dimensional temperature time series data are analyzed to determine the presence of breast tissue abnormalities. The objective of this paper is to present the scientific background of CBM and also to describe the history around the design and development of the technology.

Results: The results of using the CBM device in the initial clinical studies are also presented. Twenty four-hour long breast skin temperature circadian rhythm data was collected from 93 benign and 108 malignant female study subjects in the initial clinical studies. The predictive model developed using these datasets could differentiate benign and malignant lesions with 78% accuracy, 83.6% sensitivity and 71.5% specificity. A pilot study of 173 female study subjects is underway, in order to validate this predictive model in an independent test population.

Conclusions: The results from the initial studies indicate that the CBM may be valuable for breast health monitoring under physician supervision for confirmation of any abnormal changes, potentially prior to other methods, such as, biopsies. Studies are being conducted and planned to validate the technology and also to evaluate its ability as an adjunct breast health monitoring device for identifying abnormalities in difficult-to-diagnose dense breast tissue.

Keywords: Breast cancer; Circadian rhythm; Data mining; Machine learning; Predictive analytics; Thermal metabolomics; Wearable device.

PubMed Disclaimer

Conflict of interest statement

Declaration of Competing Interest The authors of this work declare that they have no conflicts of interest.