Isotope dilution/mass spectrometry of serum cholesterol with [3,4-13C]cholesterol: proposed definitive method
- PMID: 3301069
Isotope dilution/mass spectrometry of serum cholesterol with [3,4-13C]cholesterol: proposed definitive method
Abstract
We describe a new gas-chromatographic/mass-spectrometric (GC/MS) isotope-dilution method for determination of serum cholesterol. The method has been fully optimized and documented to provide the high accuracy and precision expected for a Definitive Method. In the presence of [3,4-13C]cholesterol, cholesteryl esters in serum are hydrolyzed under optimum conditions and the entire cholesterol pool is extracted and derivatized to silyl ethers. The cholesterol derivatives are resolved from other sterols by gas-liquid chromatography on a fused silica column, and selected ions characteristic of cholesterol and the [3,4-13C]cholesterol are monitored with a GC/MS quandrupole system. We estimated the cholesterol content of samples by bracketing each sample with standards of comparable cholesterol concentration that also contained the [3,4-13C]cholesterol. The procedure was highly reproducible (CV less than 0.5%), better accuracy and precision being obtained with [3,4-13C]cholesterol than with heptadeuterated cholesterol. Mean values per gram of dry serum for one serum pool assayed by this method and that of the National Bureau of Standards differed by 0.5%. We conclude that the method satisfies the criteria for a Definitive Method.
Similar articles
-
Total serum cholesterol by isotope dilution/mass spectrometry: a candidate definitive method.Clin Chem. 1980 Jun;26(7):854-60. Clin Chem. 1980. PMID: 6991160
-
Comparison of gas chromatography and liquid chromatography mass spectrometric measurements for high accuracy analysis of cholesterol in human serum by isotope dilution mass spectrometry.Rapid Commun Mass Spectrom. 2002;16(9):848-53. doi: 10.1002/rcm.646. Rapid Commun Mass Spectrom. 2002. PMID: 11948815
-
Determination of serum cholesterol by stable isotope dilution method using discharge-assisted thermospray liquid chromatography/mass spectrometry.Biol Mass Spectrom. 1993 Apr;22(4):247-50. doi: 10.1002/bms.1200220406. Biol Mass Spectrom. 1993. PMID: 8481413
-
Precision of glucose measurements in control sera by isotope dilution/mass spectrometry: proposed definitive method compared with a reference method.Clin Chem. 1987 Aug;33(8):1397-402. Clin Chem. 1987. PMID: 3301068
-
Chromatographic separation of cholesterol in foods.J Chromatogr. 1992 Oct 30;624(1-2):369-88. doi: 10.1016/0021-9673(92)85689-q. J Chromatogr. 1992. PMID: 1494014 Review.
Cited by
-
ACP Broad Sheet no 151: September 1997. Investigation of dyslipidaemias.J Clin Pathol. 1997 Sep;50(9):721-34. doi: 10.1136/jcp.50.9.721. J Clin Pathol. 1997. PMID: 9389972 Free PMC article. Review. No abstract available.
-
Total cholesterol performance of Abell-Levy-Brodie-Kendall reference measurement procedure: Certification of Japanese in-vitro diagnostic assay manufacturers through CDC's Cholesterol Reference Method Laboratory Network.Clin Chim Acta. 2015 May 20;445:127-32. doi: 10.1016/j.cca.2015.03.026. Epub 2015 Mar 26. Clin Chim Acta. 2015. PMID: 25818239 Free PMC article.
Publication types
MeSH terms
Substances
LinkOut - more resources
Medical
Miscellaneous