Resistance to Cyclic Fatigue of Nickel-Titanium Files Immersed in Sodium Hypochlorite at Body Temperature
- PMID: 33014060
- PMCID: PMC7520684
- DOI: 10.1155/2020/8830163
Resistance to Cyclic Fatigue of Nickel-Titanium Files Immersed in Sodium Hypochlorite at Body Temperature
Abstract
Materials and methods: Forty-five files from each brand were randomly assigned to three groups (n = 15) and subjected to the following: no immersion (control), 1-minute immersion, and 5-minute immersion in 2.5% NaOCl at 37°C. CF for all the files was tested within a well-lubricated stainless-steel artificial canal in a water bath at 37°C simulating body temperature. The procedure was video recorded, and the number of cycles to fracture (NCF) was calculated by multiplying the time taken to fracture, with the number of rotations per second. The data were analyzed for normality, and thereafter, one-way ANOVA with multiple Bonferroni was used as the post hoc test.
Results: The resistance to CF of OC files was significantly higher than PTG files in all groups. In groups immersed in NaOCl for one minute, OC files showed significant drop in the CF resistance; PTG files showed a drop as well but not significantly. Both files demonstrated a significant drop in resistance to CF after immersion in NaOCl for five minutes compared to the control group.
Conclusion: OC files were significantly more resistant to CF compared to PTG in all groups. Immersion in 2.5% NaOCl for 5 minutes significantly reduced the resistance to CF for all the files.
Copyright © 2020 Hind F. Abuhulaibah and Ammar AbuMostafa.
Conflict of interest statement
The authors declare that there are no conflicts of interest.
Similar articles
-
Cyclic Fatigue Resistance of Heat-treated Nickel-titanium Instruments after Immersion in Sodium Hypochlorite and/or Sterilization.J Endod. 2018 Apr;44(4):648-653. doi: 10.1016/j.joen.2017.12.011. Epub 2018 Feb 1. J Endod. 2018. PMID: 29397218
-
Cyclic fatigue resistance of nickel-titanium instruments after immersion in irrigant solutions with or without surfactants.J Endod. 2014 Aug;40(8):1245-9. doi: 10.1016/j.joen.2014.02.005. Epub 2014 Mar 28. J Endod. 2014. PMID: 25069942
-
Cyclic fatigue resistance of three different nickel-titanium instruments after immersion in sodium hypochlorite.J Endod. 2011 Aug;37(8):1139-42. doi: 10.1016/j.joen.2011.04.008. Epub 2011 May 31. J Endod. 2011. PMID: 21763909
-
Cyclic fatigue resistance of D-RaCe, ProTaper, and Mtwo nickel-titanium retreatment instruments after immersion in sodium hypochlorite.Clin Oral Investig. 2016 Jul;20(6):1175-9. doi: 10.1007/s00784-015-1611-4. Epub 2015 Sep 30. Clin Oral Investig. 2016. PMID: 26419673
-
Effect of Immersion in Disinfectants on Cyclic Fatigue Resistance of Nickel-Titanium Instruments: An in Vitro Study.Front Dent. 2023 May 10;20:15. doi: 10.18502/fid.v20i15.12685. eCollection 2023. Front Dent. 2023. PMID: 37312818 Free PMC article.