The influence of prosthetic crown height and implant-abutment connection design selection on the long-term implant-abutment stability: A laboratory study

Affiliations

The influence of prosthetic crown height and implant-abutment connection design selection on the long-term implant-abutment stability: A laboratory study

Aimen Bagegni et al. J Mech Behav Biomed Mater. 2021 Jan.

Abstract

Background and aim: Long-term edentulism associated with vertical loss of alveolar bone might lead to increased suprastructure height. This study aimed to evaluate the effect of suprastructure height on the stability of the implant-abutment connection by investigating the stability of two different two-piece titanium implants with internal hexagonal or conical connections under simulated oral loading conditions.

Materials and methods: A total of 48 specimens were used. The specimens were divided into 2 groups according to their implant-abutment connection (group H: internal hex connection, group C: conical connection). Each group was further divided into 3 groups according to the applied suprastructure height (H1; C1: 10 mm, H2; C2: 14 mm and H3; C3: 18 mm) (n = 8). All specimens were subjected to a cyclic loading force of 98 N for 5 million simulated chewing cycles. Then, all implants that survived the chewing simulation were quasi-statically loaded until failure. The monotonic-failure load and monotonic-bending moment at failure were evaluated.

Results: After the dynamic chewing loading, the implants showed the following survival rates: group H: 95.8%; group C: 100%. The implant suprastructures revealed survival rates of 100% and 91.5% for groups H and C, respectively. After the artificial chewing simulation of 5 million cycles, some implants in the groups with higher crowns (14 mm and 18 mm) showed crack formation and plastic deformations under the light microscope. Regarding monotonic-failure load, implants with shorter suprastructures (10 mm) revealed higher resistance to failure (C1: 1496 and H1: 1201 N) than longer suprastructures (18 mm) (C3: 465 and H3: 585 N) which was expected. The mean monotonic-bending moment values at failure ranged from 400.7 Ncm to 673.3 Ncm.

Conclusion: Implant-supported restorations with increased crown height are considered stable for an extended time period (5 million cycles which equals approximately 20 years clinical service) and a reliable treatment option in case of increased inter-arch distance. There was no difference in stability of the two internal connections. Nevertheless, the integrity of implant components might be impaired when crowns with increased heights are applied.

Keywords: Bending moment; Crack formation; Crown height; Failure resistance; Implant-abutment connection; Plastic deformation.