Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
. 2021 Feb:51:102259.
doi: 10.1016/j.msksp.2020.102259. Epub 2020 Sep 14.

Description and psychometric properties of a prototype to test tactile acuity in the neck

Affiliations

Description and psychometric properties of a prototype to test tactile acuity in the neck

Nick A Olthof et al. Musculoskelet Sci Pract. 2021 Feb.

Abstract

Background: Clinical tools assessing tactile acuity in people with persistent pain have limitations. Therefore, a novel and semi-automated tool was developed: The Imprint Tactile Acuity Device (iTAD).

Aim: To describe the iTAD prototype and present the psychometric properties of its tactile acuity assessments: the localisation test, the orientation test and the overall score (mean of both tests).

Method: A test-retest design with fifty healthy participants was used to assess intra-rater reliability (ICC(2.1)), internal consistency (ICC(2.4)) and measurement error (SEM) of the three assessments (0-100% accuracy scores each) performed at the neck. Using a known-group comparison design, balanced by age and sex, scores of thirty individuals with persistent neck pain were compared to thirty healthy controls to determine construct validity.

Results: The ICC(2,1) and ICC(2,4) were 0.60 and 0.78 for the localisation test, 0.66 and 0.77 for the orientation test, and 0.73 and 0.84 for the overall score. The SEMs were 9.0%, 8.1% and 6.0%, respectively. No fixed or proportional bias, or signs of heteroscedasticity were observed. Overall, no between group differences were observed (p = 0.49). In the male subgroup, however, the overall score was lower for people with neck pain than for healthy participants (mean difference (SE); 7.6% (2.5); p = 0.008).

Discussion: The tactile acuity assessments of the iTAD demonstrate moderate reliability and good internal consistency. Measurement errors appear comparable to currently preferred methods. Clear construct validity was not established, but results may be biased by design issues of the prototype. Taken together, the iTAD shows promise but further fine-tuning is needed.

Keywords: Chronic pain (MeSH); Neck (MeSH); Reliability (MeSH); Tactile acuity; Touch perception (MeSH); Validity (MeSH).