Clinical evaluation of an application aid for less-invasive surfactant administration (LISA)
- PMID: 33023914
- PMCID: PMC7907548
- DOI: 10.1136/archdischild-2020-319792
Clinical evaluation of an application aid for less-invasive surfactant administration (LISA)
Abstract
Background: Less-invasive surfactant administration (LISA) is increasingly used. We investigated the feasibility of a new LISA-device (Neofact®) in neonates.
Design: Prospective observational pilot study with open-label LISA in two tertiary neonatal intensive care units.
Patients: 20 infants with a gestational age of ≥26+0/7 weeks and an indication for LISA (Respiratory Severity Score (RSS)≥5 or fraction of inspired oxygen (FiO2) ≥0.30). Infants with respiratory tract malformations or unavailability of an instructed neonatologist were excluded.
Main outcome measures: Success of LISA, defined as laryngoscopy-confirmed intratracheal catheter position or a decrease in FiO2 by ≥0.05 or to 0.21, accompanied by an RSS decrease of ≥2; number of attempts needed for tracheal catheterisation.
Results: 20/57 screened infants were enrolled. Successful application occurred in 19/20 (95%). One application failed after three attempts. No device-related adverse events occurred. The median number of attempts was 2, success rate per attempt 19/31 (61%).
Conclusion: LISA via Neofact® appears feasible.
Keywords: neonatology; resuscitation; technology.
© Author(s) (or their employer(s)) 2021. Re-use permitted under CC BY-NC. No commercial re-use. See rights and permissions. Published by BMJ.
Conflict of interest statement
Competing interests: CAM has a financial relationship in the profits of the product as an inventor. The other authors have indicated they have no financial relationships relevant to this article to disclose.