A Comparative Study Between Conventional ICP-OES and the Innovative PLS Model-Assisted ICP-OES for the Assay of Trace Elements
- PMID: 33035314
- DOI: 10.1093/jaoacint/qsaa090
A Comparative Study Between Conventional ICP-OES and the Innovative PLS Model-Assisted ICP-OES for the Assay of Trace Elements
Abstract
Background: Inductively coupled plasma is widely used for elemental analysis with the advantage of being eco-friendly since the discharge is free of contaminants.
Objective: A rapid, novel method was developed for the quantitation of trace elements using inductively coupled plasma with optical emission spectrometry.This method has the advantage of simultaneous calibration compared to the conventional method.
Method: The assay was carried out for iron, copper, zinc, and molybdenum using the linear regression model partial least-squares.
Results: The method was optimized and validated as per the International Conference on Harmonization guidelines, showing highly accurate and precise results. The linearity range was 0.25-4 ppm for all trace elements under investigation. The method was applied for the assay of the cited elements in non-chelated and amino acid chelated multi-mineral preparations in the Egyptian market with acceptable mean percent recovery.
Conclusions: In comparison with the official method by flame emission, statistical analysis showed no significant difference with Student's t-test and F-values.
Highlights: Inductively coupled plasma is superior as all of the elements can be measured simultaneously. The method was found to have a high degree of specificity and can be easily applied in routine elemental analysis in laboratories.
© AOAC INTERNATIONAL 2020. All rights reserved. For permissions, please email: journals.permissions@oup.com.
Similar articles
-
Determination of macro, micro and trace elements in citrus fruits by inductively coupled plasma-optical emission spectrometry (ICP-OES), ICP-mass spectrometry and direct mercury analyzer.J Sci Food Agric. 2019 Mar 15;99(4):1870-1879. doi: 10.1002/jsfa.9382. Epub 2018 Nov 5. J Sci Food Agric. 2019. PMID: 30264403
-
Investigation of elemental composition in red teff grains using inductively coupled plasma optical emission spectroscopy (ICP OES), Sire District, Arsi zone, Ethiopia.J Trace Elem Med Biol. 2024 May;83:127389. doi: 10.1016/j.jtemb.2024.127389. Epub 2024 Jan 11. J Trace Elem Med Biol. 2024. PMID: 38242004
-
Low gas flow inductively coupled plasma optical emission spectrometry for the analysis of food samples after microwave digestion.Talanta. 2014 Nov;129:575-8. doi: 10.1016/j.talanta.2014.06.045. Epub 2014 Jun 28. Talanta. 2014. PMID: 25127635
-
How ICP-OES changed the face of trace element analysis: Review of the global application landscape.Sci Total Environ. 2023 Dec 20;905:167242. doi: 10.1016/j.scitotenv.2023.167242. Epub 2023 Sep 22. Sci Total Environ. 2023. PMID: 37742982 Review.
-
The role of copper, molybdenum, selenium, and zinc in nutrition and health.Clin Lab Med. 1998 Dec;18(4):673-85. Clin Lab Med. 1998. PMID: 9891606 Review.
MeSH terms
Substances
LinkOut - more resources
Full Text Sources