. 2021 Oct;35(10):921-926.
doi: 10.1038/s41371-020-00424-0. Epub 2020 Oct 9.

Results of a project to calibrate mercury sphygmomanometer blood pressure-measuring devices in Egypt

Affiliations

Results of a project to calibrate mercury sphygmomanometer blood pressure-measuring devices in Egypt

Soliman Ghareeb et al. J Hum Hypertens. 2021 Oct.

Abstract

Hypertension (HTN) is a common health problem and a major cardiovascular risk factor. Accurate measurement of blood pressure (BP) is mandatory for proper diagnosis and follow-up. The aim of this study was to evaluate the accuracy of mercury sphygmomanometer BP-measuring devices in public hospitals in Cairo, Egypt. Fifty public hospitals were included, and 10% of all mercury sphygmomanometer devices in each hospital were tested. Assessment included physical condition (e.g., mercury status, lid of the device, state of the rubber tubes), leakage rate, and calibration accuracy (as compared with a reference device). Devices were approved as accurate when they could successfully pass all three assessment tests. The total number of sphygmomanometer devices was 465. The overall pass rate was 1.3% (six devices). Twenty-five (5.2%) devices passed all of the physical tests, 50 (10.8%) passed the leakage test, and 50 (16.5%) passed the calibration accuracy test. There were 162 (34.8%) devices that showed a high leakage rate (>80 mmHg) and thus were not tested for calibration accuracy. In conclusion, most of the mercury sphygmomanometer devices in hospitals are neglected and not checked regularly for any errors. A plan should be made to gradually replace those failed devices with new, validated, and well-calibrated devices, preferably devices that do not contain mercury.

PubMed Disclaimer

Conflict of interest statement

The authors declare that they have no conflict of interest.

Figures

Fig. 1
Fig. 1
Failure rate at the different check points in the physical assessment of the devices.
Fig. 2
Fig. 2
Frequency of the different reasons of failed mercury check in the tested devices.
Fig. 3
Fig. 3
Frequency of the different reasons of failed health and safety check in the tested devices.

References

    1. Benjamin EJ, Virani SS, Callaway CW, Chamberlain AM, Chang AR, Cheng S, et al. Heart Disease and Stroke Statistics—2018 update: a report from the American Heart Association. Circulation. 2018;137:e67–e492.. doi: 10.1161/CIR.0000000000000558. - DOI - PubMed
    1. World Health Organization Roft EM. New round of STEPwise survey 2016–2017 to start in Egypt. 2017. http://www.emro.who.int/egy/egypt-infocus/who-will-start-a-new-round-of-....
    1. Ibrahim MM, Rizk H, Appel LJ, el Aroussy W, Helmy S, Sharaf Y, NHP Investigative Team et al. Hypertension prevalence, awareness, treatment, and control in Egypt. Results from the Egyptian National Hypertension Project (NHP). Hypertension. 1995;26:886–90. doi: 10.1161/01.HYP.26.6.886. - DOI - PubMed
    1. WHO. STEPwise approach to chronic disease risk factor surveillance: World Health Organization. 2017. https://www.who.int/ncds/surveillance/steps/egypt/en/.
    1. Forouzanfar MH, Liu P, Roth GA, Ng M, Biryukov S, Marczak L, et al. Global burden of hypertension and systolic blood pressure of at least 110 to 115 mmHg, 1990–2015. JAMA. 2017;317:165–82. doi: 10.1001/jama.2016.19043. - DOI - PubMed

Publication types