. 2021 Jan 4;59(1):199-206.
doi: 10.1093/ejcts/ezaa256.

Safety and performance of a novel transventricular beating heart mitral valve repair system: 1-year outcomes

Affiliations

Safety and performance of a novel transventricular beating heart mitral valve repair system: 1-year outcomes

James S Gammie et al. Eur J Cardiothorac Surg. .

Abstract

Objectives: The objective of this study was to evaluate the safety and performance of a novel, beating heart procedure that enables echocardiographic-guided beating heart implantation of expanded polytetrafluoroethylene (ePTFE) artificial cords on the posterior mitral leaflet of patients with degenerative mitral regurgitation.

Methods: Two prospective multicentre studies enrolled 13 (first-in-human) and 52 subjects, respectively. Patients were treated with the HARPOON beating heart mitral valve repair system. The primary (30-day) end point was successful implantation of cord(s) with mitral regurgitation reduction to ≤moderate. An independent core laboratory analysed echocardiograms.

Results: Of 65 patients enrolled, 62 (95%) achieved technical success, 2 patients required conversion to open surgery and 1 procedure was terminated. The primary end point was met in 59/65 (91%) patients. Among the 62 treated patients, the mean procedural time was 2.1 ± 0.5 h. Through discharge, there were no deaths, strokes or renal failure events. At 1 year, 2 of the 62 patients died (3%) and 8 (13%) others required reoperations. At 1 year, 98% of the patients with HARPOON cords were in New York Heart Association class I or II, and mitral regurgitation was none/trace in 52% (n = 27), mild in 23% (n = 12), moderate in 23% (n = 12) and severe in 2% (n = 1). Favourable cardiac remodelling outcomes at 1 year included decreased end-diastolic left ventricular volume (153 ± 41 to 119 ± 28 ml) and diameter (53 ± 5 to 47 ± 6 mm), and the mean transmitral gradient was 1.4 ± 0.7 mmHg.

Conclusions: This initial clinical experience with the HARPOON beating heart mitral valve repair system demonstrates encouraging early safety and performance.

Clinical registration numbers: NCT02432196 and NCT02768870.

Keywords: Echocardiography; Mitral regurgitation; Surgery; Valvuloplasty.