Pacemaker-mediated tachycardia in a dual-lead CRT-D: What is the mechanism?
- PMID: 33058215
- DOI: 10.1111/pace.14089
Pacemaker-mediated tachycardia in a dual-lead CRT-D: What is the mechanism?
Abstract
A 73-year-old gentleman with dilated cardiomyopathy, left bundle branch block and a left ventricular (LV) ejection fraction of 20% was implanted with two LV leads in a tri-ventricular cardiac resynchronisation therapy defibrillator (CRT-D) trial. As a part of the trial he was programmed with fusion-based CRT therapy with dual LV lead only pacing. The patient presented to local heart failure service 12 years after implant, after a positive response to CRT therapy, with increase in fatigue, shortness of breath and bilateral pitting oedema. The patient sent a remote monitoring transmission that suggested loss of capture on one of the LV leads. This coupled with atrial ectopics was producing a high burden of pacemaker-mediated tachycardia (PMT) that was not seen when both LV leads had been capturing. What is the mechanism for this? Dual LV-lead tri-ventricular leads have been shown to have variable improvements in CRT response but with an increased complexity of implant procedure. This is the first case report of PMT-induced heart failure exacerbation in a tri-ventricular device following loss of LV capture of one lead.
Keywords: TriVentricular; cardiac resynchronisation therapy; dual left ventricular lead; heart failure exacerbation; pacemaker-mediated tachycardia.
© 2020 Wiley Periodicals LLC.
References
REFERENCES
-
- Rogers DP, Lambiase PD, Lowe MD, Chow AWC. A randomized double-blind crossover trial of triventricular versus biventricular pacing in heart failure. Eur J Heart Fail. 2012;14:495-505.
-
- Zhang B, Guo J, Zhang G. Comparison of triple-site ventricular pacing versus conventional cardiac resynchronization therapy in patients with systolic heart failure: a meta-analysis of randomized and observational studies. J Arrhythm. 2018;34:55-64.
-
- Lenarczyk R, Kowalski O, Sredniawa B, et al. Implantation feasibility, procedure-related adverse events and lead performance during 1-year follow-up in patients undergoing triple-site cardiac resynchronization therapy: a substudy of TRUST CRT randomized trial. J Cardiovasc Electrophysiol. 2012;23(11):1228-1236.
-
- Ausubel K, Gabry MD, Klementowicz PT, Furman S. Pacemaker-mediated endless loop tachycardia at rates below the upper rate limit. Am J Cardiol. 1988;61:465-467.
-
- Barold SS. Repetitive reentrant and non-reentrant ventriculoatrial synchrony in dual chamber pacing. Clin Cardiol. 1991;14:754-763.