Utility of an Automatic Vision-Based Examination System (AVE-562) for the Detection of Clonorchis sinensis Eggs in Stool
- PMID: 33063684
- PMCID: PMC7591289
- DOI: 10.3343/alm.2021.41.2.221
Utility of an Automatic Vision-Based Examination System (AVE-562) for the Detection of Clonorchis sinensis Eggs in Stool
Abstract
Stool examination is the gold standard for the detection of intestinal parasites. We assessed the performance of a newly developed AVE-562 analyzer (AVE Science & Technology Co., Hunan, China) for the vision-based detection of eggs of Clonorchis sinensis-the most common intestinal parasite in Korea-in stool samples. In total, 30 stool samples with a high or low egg count or without eggs (as negative control samples) (N=10 each) were prepared and analyzed. The performance of the AVE-562 analyzer was compared with that of the formalin-ether concentration (FEC) method. The overall correct identification rate of the AVE-562 analyzer based on FEC results was 66.6%. The sensitivity, specificity, positive predictive value, and negative predictive value of the AVE-562 analyzer for detecting C. sinensis eggs were 36.4%, 100.0%, 100.0%, and 73.1%, respectively. The average time required to run five tests simultaneously was 27 min using the AVE-562 analyzer and 58 min using the FEC method. Although the AVE-562 analyzer enables rapid and convenient stool examination, its sensitivity needs to be improved, particularly considering the prevalence of low-burden C. sinensis infection in Korea.
Keywords: AVE-562 analyzer; Clonorchis sinensis; Eggs; Parasite; Performance.
Conflict of interest statement
No potential conflicts of interest relevant to this article were reported.
Figures
References
-
- Korea Centers for Disease Control and Prevention; Korea National Institute of Health. National survey of the prevalence of intestinal parasitic infections in Korea, 2012. The 8th Report. Osong. Chungcheongbuk-do, Korea: 2013.
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources