Comparative Study
. 2020 Oct 15;30(3):030704.
doi: 10.11613/BM.2020.030704.

Test results comparison and sample stability study: is the BD Barricor tube a suitable replacement for the BD RST tube?

Affiliations
Comparative Study

Test results comparison and sample stability study: is the BD Barricor tube a suitable replacement for the BD RST tube?

Sanja Mandić et al. Biochem Med (Zagreb). .

Abstract

Introduction: The aim was to evaluate the BD Barricor tubes by comparison with the BD Rapid Serum Tubes (RST) through measuring 25 analytes and monitoring sample stability after 24 hours and 7 days.

Materials and methods: Samples of 52 patients from different hospital departments were examined. Blood was collected in BD RST and BD Barricor tubes (Becton, Dickinson and Company, Franklin Lakes, USA). Analytes were measured by Beckman Coulter AU 480 (Beckman Coulter, Brea, USA), Dimension EXL (Siemens Healthcare Diagnostics, Newark, USA) and ARCHITECT i2000SR (Abbott Diagnostics, Lake Forest, USA). Between-tube comparison for each analyte was performed, along with testing analyte stability after storing samples at 4 °C.

Results: BD Barricor tubes showed unacceptable bias compared to BD RST tubes for potassium (K) (- 4.5%) and total protein (4.4%). Analyte stability after 24 hours was acceptable in both tested tubes for most of analytes, except for glucose, aspartate aminotransferase (AST) and lactate dehydrogenase (LD) in BD Barricor and free triiodothyronine in BD RST sample tubes. Analyte stability after 7 days was unacceptable for sodium, K, calcium, creatine kinase isoenzyme MB, AST, LD and troponin I in both samples; additionally for glucose, alkaline phosphatase and albumin in BD Barricor.

Conclusion: All analytes, except K and total protein, can be measured interchangeably in BD RST and BD Barricor tubes, applying the same reference intervals. For most of the analytes, sample re-analysis can be performed in both tubes after 24 hours and 7 days, although BD RST tubes show better 7-day analytes stability over BD Barricor tubes.

Keywords: BD Barricor tube; evaluation; plasma; serum; stability.

PubMed Disclaimer

Conflict of interest statement

Potential conflict of interest None declared.

References

    1. Lippi G, Cornes MP, Grankvist K, Nybo M, Simundic AM, Working Group for Preanalytical P et al. EFLM WG-Preanalytical phase opinion paper: local validation of blood collection tubes in clinical laboratories. Clin Chem Lab Med. 2016;54:755–60. 10.1515/cclm-2015-1274 - DOI - PubMed
    1. Clinical and Laboratory Standards Institute (CLSI). CLSI document GP34-A. Validation and Verification of Tubes for Venous and Capillary Blood Specimen Collection; Approved Guideline. Wayne, PA: CLSI; 2010
    1. Guder WG, Narayanan S. Pre-examination procedures in laboratory diagnostics: preanalytical aspects and their impact on the quality of medical laboratory results. Berlin: De Gruyter, 2015. https://doi.org/10.1515/9783110334043 - DOI
    1. Balbás LA, Amaro MS, Rioja RG, Martin MJ, Soto AB. Stability of plasma electrolytes in Barricor and PST II tubes under different storage conditions. Biochem Med (Zagreb). 2017;27:225–30. 10.11613/BM.2017.024 - DOI - PMC - PubMed
    1. Kocijancic M, Cargonja J, Delic-Knezevic A. Evaluation of the BD Vacutainer RST blood collection tube for routine chemistry analytes: clinical significance of differences and stability study. Biochem Med (Zagreb). 2014;24:368–75. 10.11613/BM.2014.039 - DOI - PMC - PubMed

Publication types