Interpretive discrepancies caused by target values inter-batch variations in chemiluminescence immunoassay for SARS-CoV-2 IgM/IgG by MAGLUMI

Affiliations

Interpretive discrepancies caused by target values inter-batch variations in chemiluminescence immunoassay for SARS-CoV-2 IgM/IgG by MAGLUMI

Iveta Selingerova et al. J Med Virol. 2021 Mar.

Abstract

Plasma specimens from coronavirus disease 2019 patients were double-tested for anti-severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) antibodies by two different batches of MAGLUMI 2019-nCov immunoglobulin M/immunoglobulin G (IgM/IgG) assays to evaluate IgM/IgG levels, qualitative interpretation, antibody kinetics, and linearity of diluted specimen. Here we show that (i) high-level IgM specimens need to be diluted with negative human plasma but not kit diluents and (ii) measured anti-SARS-CoV-2 IgM/IgG concentrations are substantially higher with later marketed immunoassay batch leading to (iii) the change of qualitative interpretation (positive vs. negative) in 12.3% of specimens measured for IgM, (iv) the informative time-course pattern of antibody production only when data from different immunoassay batches are not combined.

Keywords: CLIA; anti-SARS-CoV-2 IgG; anti-SARS-CoV-2 IgM; inter-assay batch variation; interpretative inconsistency.

PubMed Disclaimer

References

REFERENCES

    1. Padoan A, Cosma C, Sciacovelli L, Faggian D, Plebani M. Analytical performances of a chemiluminescence immunoassay for SARS-CoV-2 IgM/IgG and antibody kinetics. Clin Chem Lab Med. 2020;58(7):1081-1088. https://doi.org/10.1515/cclm-2020-0443
    1. Infantino M, Grossi V, Lari B, et al. Diagnostic accuracy of an automated chemiluminescent immunoassay for anti-SARS-CoV-2 IgM and IgG antibodies: an Italian experience. J Med Virol. 2020;92(9):1671-1675. https://doi.org/10.1002/jmv.25932
    1. Lippi G, Salvagno GL, Pegoraro M, et al. Assessment of immune response to SARS-CoV-2 with fully automated MAGLUMI 2019-nCoV IgG and IgM chemiluminescence immunoassays. Clin Chem Lab Med. 2020;58(7):1156-1159. https://doi.org/10.1515/cclm-2020-0473
    1. Montesinos I, Gruson D, Kabamba B, et al. Evaluation of two automated and three rapid lateral flow immunoassays for the detection of anti-SARS-CoV-2 antibodies. J Clin Virol. 2020;128:104413. https://doi.org/10.1016/j.jcv.2020.104413
    1. Plebani M, Padoan A, Negrini D, Carpinteri B, Sciacovelli L. Diagnostic performances and thresholds: the key to harmonization in serological SARS-CoV-2 assays? Clin Chim Acta. 2020;509:1-7. https://doi.org/10.1016/j.cca.2020.05.050

Publication types

LinkOut - more resources