Fracture of an S-ICD lead after two prior transvenous lead-related complications with conventional defibrillators
- PMID: 33085093
- DOI: 10.1111/pace.14101
Fracture of an S-ICD lead after two prior transvenous lead-related complications with conventional defibrillators
Abstract
Six months after subcutaneous implantable cardioverter defibrillator (S-ICD) implantation a 26-year-old Brugada patient presented because of a beeping tone emitted by his device. Chest X-ray displayed two functionless transvenous shock leads and the S-ICD system with a lead fracture. During lead revision procedure, extensive preparation of the lead from unexpectedly firm surrounding fibrous tissue encapsulating the lead was necessary before it could be removed, and a new shock lead could be implanted. This is the first report of an S-ICD lead exchange due to very early lead fracture and unexpectedly severe fibrous tissue hampering surgical lead extraction.
Keywords: implantable defibrillator; lead failure; subcutaneous defibrillator.
© 2020 The Authors. Pacing and Clinical Electrophysiology published by Wiley Periodicals LLC.
References
REFERENCES
-
- Priori S, Blomström-Lundqvist C, Mazzanti A, et al. ESC guidelines for the management of patients with ventricular arrhythmias and the prevention of sudden cardiac death: the task force for the management of patients with ventricular arrhythmias and the prevention of sudden cardiac death of the European Society of Cardiology (ESC). Endorsed by: Association for European Paediatric and Congenital Cardiology (AEPC). Eur Heart J. 2015;36:2793-2867.
-
- Kirkfeldt RE, Johansen JB, Nohr EA, Jørgensen OD, Nielsen JC. Complications after cardiac implantable electronic device implantations: an analysis of a complete, nationwide cohort in Denmark. Eur Heart J. 2014;35:1186-1194.
-
- Kleemann T, Nonnenmacher F, Strauss M, et al. Long-term performance and lead failure analysis of the durata defibrillation lead compared to its previous model, the recalled Riata defibrillation lead. J Cardiovasc Electrophysiol. 2019;30:2012-2019.
-
- Baalman SWE, Quast ABE, Brouwer TF, Knops RE. An overview of clinical outcomes in transvenous and subcutaneous ICD patients. Curr Cardiol Rep. 2018;20:72.
-
- Olde Nordkamp LR, Knops RE, Bardy GH, et al. Rationale and design of the PRAETORIAN trial: a Prospective, RAndomizEd comparison of subcuTaneOus and tRansvenous ImplANtable cardioverter-defibrillator therapy. Am Heart J. 2012;163:753-760.