Comparative accuracy of cone-beam CT and conventional multislice computed tomography for real-time navigation in zygomatic implant surgery
- PMID: 33112508
- DOI: 10.1111/cid.12958
Comparative accuracy of cone-beam CT and conventional multislice computed tomography for real-time navigation in zygomatic implant surgery
Abstract
Background: Cone-beam computed tomography (CBCT) and conventional multislice CT (MSCT) are both used in zygomatic implant navigation surgery but the superiority of one technique versus the other remains unclear.
Purpose: This study compared the accuracy of CBCT and MSCT in zygomatic implant navigation surgery by calculating the deviations of implants.
Material and methods: Patients with severely atrophic maxillae were classified into two groups according to the use of CBCT- or MSCT-guided navigation system. The entry and apical distance deviation, and the angle deviation of zygomatic implants were measured on fused operation images. A linear effect model was used for analysis, with statistical significance set at P < .05.
Results: A total of 72 zygomatic implants were inserted as planned in 23 patients. The comparison of deviations in CBCT and MSCT groups showed a mean (± SD) entry deviation of 1.69 ± 0.59 mm vs 2.04 ± 0.78 mm (P = .146), apical deviation of 2 ± 0.68 mm vs 2.55 ± 0.85 (P < .001), and angle deviation of 2.32 ± 1.02° vs 3.23 ± 1.21° (P = .038).
Conclusion: Real-time zygomatic implant navigation surgery with CBCT may result in higher values for accuracy than MSCT.
Keywords: accuracy; computer-assisted surgery; edentulous maxillae; real-time navigation surgery; zygomatic implants.
© 2020 Wiley Periodicals LLC.
References
REFERENCES
-
- Chrcanovic BR, Pedrosa AR, Neto Custodio AL. Zygomatic implants: a critical review of the surgical techniques. Oral Maxillofac Surg. 2013;17:1-9.
-
- Schramm A, Gellrich NC, Schimming R, Schmelzeisen R. Computer-assisted insertion of zygomatic implants (Branemark system) after extensive tumor surgery. [Article in German]. Mund Kiefer Gesichtschir. 2000;4(5):292-295.
-
- Xiaojun C, Yanping L, Yiqun W, Chengtao W. Computer-aided oral implantology: methods and applications. J Med Eng Technol. 2007;31(6):459-467.
-
- Wu Y, Wang F, Huang W, Fan S. Real-time navigation in zygomatic implant placement: workflow. Oral Maxillofac Surg Clin North Am. 2019;31(3):357-367.
-
- Vercruyssen M, Fortin T, Widmann G, Jacobs R, Quirynen M. Different techniques of static/dynamic guided implant surgery: modalities and indications. Periodontol 2000. 2014;66(1):214-227.
MeSH terms
Substances
Grants and funding
- 2017-1-005/Ninth People's Hospital affiliated to Shanghai Jiao Tong University, School of Medicine "Multi-Disciplinary Team" Clinical Research Project
- 2017YFB1302904/Science and Technology Ministry of China
- the Integrated Fund Project of Ninth People's Hospital
- 81700944/Youth Program of National Natural Science Foundation of China
LinkOut - more resources