Comparative Study
. 1987 Nov;33(11):2096-100.

Chemiluminescence immunoassay of thyrotropin with acridinium-ester-labeled antibody evaluated and compared with two other immunoassays

Affiliations
  • PMID: 3315305
Comparative Study

Chemiluminescence immunoassay of thyrotropin with acridinium-ester-labeled antibody evaluated and compared with two other immunoassays

M P Bounaud et al. Clin Chem. 1987 Nov.

Abstract

A new chemiluminometric immunoassay of thyrotropin (TSH) involves antibody labeled with acridinium ester ("Magic Lite System," Ciba Corning Diagnostic Corp.). The assay is rapid, with two incubations totaling 2.5 h, requires two standards per run, and takes 10 s per sample for the quantification step. Analytical performance, within- and between-run reproducibilities, and linearity were excellent. The detection limit is 0.04 milli-int. unit/L. Results correlated well with those obtained by immunoradiometric assay (RIA-gnost hTSH, Hoechst-Behring) and immunofluorometric assay (hTSH Delfia, LKB): r = 0.975. TSH measurements in 32 euthyroid subjects ranged from 0.4 to 4.8 milli-int. units/L (mean 1.35 milli-int. units/L). TSH values for 51 hypothyroid and subclinically hypothyroid patients ranged from 2 to 65 milli-int. units/L. TSH values for 33 hyperthyroid patients (less than 0.14 milli-int. unit/L, less than 0.04 milli-int. unit/L in 16 of the 33) were clearly lower than for most untreated euthyroid subjects. For 169 other individuals whose thyroid function was being routinely assessed. TSH ranged from 0.4 to 4.8 milli-int. units/L, three had TSH less than 0.14 milli-int. unit/L, and four had TSH between 0.14 and 0.4 milli-int. unit/L. This system is as efficient and reliable for screening for thyroid function as the two comparison systems.