Comparison of two real-time polymerase chain reaction assays for the detection of severe acute respiratory syndrome-CoV-2 from combined nasopharyngeal-throat swabs
- PMID: 33154251
- PMCID: PMC7709598
- DOI: 10.4103/ijmm.IJMM_20_279
Comparison of two real-time polymerase chain reaction assays for the detection of severe acute respiratory syndrome-CoV-2 from combined nasopharyngeal-throat swabs
Abstract
Context: In the absence of effective treatment or vaccine, the current strategy for the prevention of further transmission of severe acute respiratory syndrome (SARS) CoV-2 (COVID-19) infection is early diagnosis and isolation of cases. The diagnosis of SARS-CoV-2 is done by detecting viral RNA in the nasopharyngeal and throat swabs by real-time polymerase chain reaction (PCR). Many commercial assays are now available for performing the PCR assay.
Aims: The aim was to evaluate the performance of the SD Biosensor nCoV real-time detection kit with the real-time PCR kit provided by the Indian Council of Medical Research-National Institute of Virology (ICMR-NIV), Pune (NIV Protocol).
Subjects and methods: A total of 253 pairs of nasopharyngeal-oropharyngeal swabs combined in a single viral transport medium were tested for viral RNA by both the protocols. The sensitivity and specificity of the SD Biosensor were calculated considering the ICMR-NIV kit as the gold standard. Matched pairs of recorded cycle threshold values (Ct values) were compared by Pearson's correlation coefficient.
Results: Concordant COVID-19 negative and positive PCR results were reported for 113 and 77 samples, respectively. The SD Biosensor kit additionally detected 62 cases, which were found negative by the NIV protocol. In all discordant positive results by the SD Biosensor kit, the average Ct values were higher than the concordant positive results. A total of forty samples tested positive for E gene by SD Biosensor and having Ct values <25 had 100% concordance with NIV protocol results and 39 samples tested positive for E gene by SD Biosensor having Ct value >32 were all found negative by the NIV protocol.
Conclusions: The results highlight the need for careful evaluation of commercial kits before being deployed for screening of COVID-19 infections.
Keywords: COVID-19; molecular diagnostics; real-time polymerase chain reaction; severe acute respiratory syndrome-CoV-2; test comparison.
Conflict of interest statement
None
Similar articles
-
Automated SARS-COV-2 RNA extraction from patient nasopharyngeal samples using a modified DNA extraction kit for high throughput testing.Ann Saudi Med. 2020 Sep-Oct;40(5):373-381. doi: 10.5144/0256-4947.2020.373. Epub 2020 Oct 1. Ann Saudi Med. 2020. PMID: 32954791 Free PMC article.
-
Analysis of multi-sample pools in the detection of SARS-CoV-2 RNA for mass screening: An Indian perspective.Indian J Med Microbiol. 2020 Jul-Dec;38(3 & 4):451-456. doi: 10.4103/ijmm.IJMM_20_273. Indian J Med Microbiol. 2020. PMID: 33154262 Free PMC article.
-
Comparison of the diagnostic efficacy between two PCR test kits for SARS-CoV-2 nucleic acid detection.J Clin Lab Anal. 2020 Oct;34(10):e23554. doi: 10.1002/jcla.23554. Epub 2020 Sep 25. J Clin Lab Anal. 2020. PMID: 32977349 Free PMC article.
-
The genetic sequence, origin, and diagnosis of SARS-CoV-2.Eur J Clin Microbiol Infect Dis. 2020 Sep;39(9):1629-1635. doi: 10.1007/s10096-020-03899-4. Epub 2020 Apr 24. Eur J Clin Microbiol Infect Dis. 2020. PMID: 32333222 Free PMC article. Review.
-
[SARS-CoV-2 and Microbiological Diagnostic Dynamics in COVID-19 Pandemic].Mikrobiyol Bul. 2020 Jul;54(3):497-509. doi: 10.5578/mb.69839. Mikrobiyol Bul. 2020. PMID: 32755524 Review. Turkish.
Cited by
-
Utility of Antigen-Based Rapid Diagnostic Test for Detection of SARS-CoV-2 Virus in Routine Hospital Settings.Lab Med. 2021 Nov 2;52(6):e154-e158. doi: 10.1093/labmed/lmab033. Lab Med. 2021. PMID: 33928384 Free PMC article.
-
Ophthalmic Emergencies and Effects of Different Quarantine Models During the COVID-19 Pandemic.Med Sci Monit. 2021 May 13;27:e931967. doi: 10.12659/MSM.931967. Med Sci Monit. 2021. PMID: 33980807 Free PMC article.
-
A Review on SERS-Based Detection of Human Virus Infections: Influenza and Coronavirus.Biosensors (Basel). 2021 Feb 28;11(3):66. doi: 10.3390/bios11030066. Biosensors (Basel). 2021. PMID: 33670852 Free PMC article. Review.
References
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Medical
Miscellaneous