Factors Associated with Heating Efficiency in Transcranial Focused Ultrasound Therapy

Affiliations

Factors Associated with Heating Efficiency in Transcranial Focused Ultrasound Therapy

Kazuaki Yamamoto et al. Neurol Med Chir (Tokyo). .

Abstract

Transcranial magnetic resonance-guided focused ultrasound (FUS) therapy is a less invasive stereotactic treatment for tremor and other movement disorders. A sufficiently high temperature in the target brain tissue is crucial during ablation procedures for good outcomes. Therefore, maximizing the heating efficiency is critical in cases where high temperature cannot be achieved because of patient-related characteristics. However, a strategy to achieve the desired therapeutic temperature with FUS has not yet been established. This study aimed to investigate the procedural factors associated with heating efficiency in FUS.We retrospectively reviewed and analyzed data from patients who underwent FUS for ventralis intermedius (VIM) nucleus thalamotomy. In all, 30 consecutive patients were enrolled. 18 with essential tremor (ET), 11 with tremor-dominant Parkinson's disease (TDPD), and 1 with Holmes tremor. Multivariate regression analysis showed that decline in heating efficiency was associated with lower skull density ratio (SDR) and a greater subtotal rise in temperature until the previous sonication. To maximize heating efficiency, the temperature increase should be set to the least value in the target alignment and verification phases, and subsequently should be increased sufficiently in the treatment phase. This strategy may be particularly beneficial in cases where high ablation temperatures cannot be achieved because of patient-related characteristics. Importantly, a broad patient population would benefit from this strategy as it could reduce the need for high energy to achieve therapeutic temperatures, thereby decreasing the risks of adverse events.

Keywords: focused ultrasound therapy; heating efficiency; ventralis intermedius nucleus.

PubMed Disclaimer

Conflict of interest statement

Conflicts of Interest Disclosure

All authors have no conflict of interest. All JNS member authors have registered online Self-reported COI Disclosure Statement Forms through the website.

Figures

Fig. 1
Fig. 1
Subtotal temperature rise and heating efficiency. The graphs show the association between the subtotal amount of temperature rise until the previous sonication and heating efficiency. (A) Heating efficiency declined as the subtotal amount of temperature rise until the previous sonication increased. (B) Similar findings were observed in both groups separated according to skull density ratios.

References

    1. Bauer R, Martin E, Haegele-Link S, Kaegi G, von Specht M, Werner B: Noninvasive functional neurosurgery using transcranial MR imaging-guided focused ultrasound. Parkinsonism Relat Disord 20: S197–S199, 2014 - PubMed
    1. Dobrakowski PP, Machowska-Majchrzak AK, Labuz-Roszak B, Majchrzak KG, Kluczewska E, Pierzchała KB: MR-guided focused ultrasound: a new generation treatment of Parkinson's disease, essential tremor and neuropathic pain. Interv Neuroradiol 20: 275–282, 2014 - PMC - PubMed
    1. Martin E, Jeanmonod D, Morel A, Zadicario E, Werner B: High-intensity focused ultrasound for noninvasive functional neurosurgery. Ann Neurol 66: 858–861, 2009 - PubMed
    1. Bond AE, Shah BB, Huss DS, et al. : Safety and efficacy of focused ultrasound thalamotomy for patients with medication-refractory, tremor-dominant parkinson disease: a randomized clinical trial. JAMA Neurol 74: 1412–1418, 2017 - PMC - PubMed
    1. Ghanouni P, Pauly KB, Elias WJ, et al. : Transcranial MRI-guided focused ultrasound: a review of the technologic and neurologic applications. Am J Roentgenol 205: 150–159, 2015 - PMC - PubMed