Reliability of the portable gait rhythmogram in post-stroke patients
- PMID: 33164920
- DOI: 10.3233/BME-206007
Reliability of the portable gait rhythmogram in post-stroke patients
Abstract
Background: Gait analysis, such as portable gait rhythmogram (PGR), provides objective information that helps in the quantitative evaluation of human locomotion.
Objective: The purpose of this study was to assess the reliability of PGR in post-stroke patients.
Methods: Two raters (A and B) examined 48 post-stroke patients. To assess intra-rater reliability, rater A tested subjects on three separate occasions (Days 1, 2, and 3). To assess inter-rater reliability, raters A and B independently tested participants on the same occasion (Day 3).
Results: There was no significant systematic bias between test occasions or raters. Intraclass correlation coefficient values were 0.93-0.97 for intra-rater reliability at both the comfortable speed and maximum speed, and 0.97-0.98 (comfortable speed) and 0.97-0.99 (maximum speed) for inter-rater reliability. The standard error was 1.25-1.49 (comfortable speed) and 1.62-1.77 (maximum speed) for intra-rater investigation, and 1.04-1.32 (comfortable speed) and 0.91-1.26 (maximum speed) for inter-rater investigation. At the 90% confidence level, the minimum detectable change ranged from 2.9-4.1%, and the error of an individual's score at a given time point ranged from ±2.1-2.9%.
Conclusions: Based on this excellent reliability of the PGR in post-stroke patients, it can be recommended as a simple test of gait analysis in this population.
Keywords: Acceleration; gait analysis; portable gait rhythmogram; reliability; stroke.
Similar articles
-
Intra-rater and inter-rater reliability of the portable gait rhythmogram in post-stroke patients.J Phys Ther Sci. 2017 May;29(5):874-879. doi: 10.1589/jpts.29.874. Epub 2017 May 16. J Phys Ther Sci. 2017. Retraction in: J Phys Ther Sci. 2017 Oct;29(10):r1. doi: 10.1589/jpts.29.874r1. PMID: 28603363 Free PMC article. Retracted.
-
Feasibility and inter-rater reliability of physical performance measures in acutely admitted older medical patients.PLoS One. 2015 Feb 23;10(2):e0118248. doi: 10.1371/journal.pone.0118248. eCollection 2015. PLoS One. 2015. PMID: 25706553 Free PMC article.
-
Inter- and intra-rater reliability of the GAITRite system among individuals with sub-acute stroke.Gait Posture. 2014;40(1):259-61. doi: 10.1016/j.gaitpost.2014.02.007. Epub 2014 Feb 26. Gait Posture. 2014. PMID: 24630463 Clinical Trial.
-
Reliability of the six-minute walk test in individuals with stroke: systematic review and meta-analysis.Neurol Sci. 2021 Jan;42(1):81-87. doi: 10.1007/s10072-020-04829-0. Epub 2020 Oct 16. Neurol Sci. 2021. PMID: 33064231
-
Intraclass correlations: uses in assessing rater reliability.Psychol Bull. 1979 Mar;86(2):420-8. doi: 10.1037//0033-2909.86.2.420. Psychol Bull. 1979. PMID: 18839484 Review.
MeSH terms
LinkOut - more resources
Full Text Sources
Medical
Research Materials