Investigation of the Bragg peak degradation caused by homogeneous and heterogeneous lung tissue substitutes: proton beam experiments and comparison to current clinical dose calculation
- PMID: 33171454
- DOI: 10.1088/1361-6560/abc938
Investigation of the Bragg peak degradation caused by homogeneous and heterogeneous lung tissue substitutes: proton beam experiments and comparison to current clinical dose calculation
Abstract
Submillimetre structures of lung tissue are not represented in computed tomography images used for radiotherapeutic dose calculation. In order to study the effect experimentally, lung substitutes with properties similar to lung tissue were chosen, namely two types of commercial lung tissue equivalent plates (LTEPs) (CIRS, USA), two types of cork, balsawood, floral foam and konjac sponge. Laterally integrated dose profiles were measured as a function of depth for proton pencil beams (PBs) with an initial nominal energy of 97.4 and 148.2 MeV, respectively. The obtained dose profiles were investigated for their shifting and degradation of the Bragg peak (BP) caused by the materials, expressed as water equivalent thickness (WET) and full width half maximum. The set-up was simulated in the treatment planning system (TPS) RayStation using the Monte Carlo (MC) dose calculation algorithm. While the WET between experiment and dose calculation agreed within 0.5 mm, except for floral foam, the full width half maximum was underestimated in the TPS by up to 2.3 mm. Normalisation to the same mass thickness of the lung substitutes allowed to classify LTEPs and balsawood as homogeneous and cork, floral foam and konjac sponge as heterogeneous materials. The material specific BP degradation was up to 3.4 times higher for the heterogeneous samples. The modulation power as a measure for the heterogeneity was compared to the spectrum of Hounsfield units (HU) of the materials. A clear correlation was not found, but with further improvements the HU spectrum may serve as an indicator for the material heterogeneity. Further, MC simulations of binary voxel models using GATE/Geant4 were performed to investigate the influence of grain size and mass density. For mass densities similar to lung tissue the BP degradation had a maximum at 3 and 7 mm grain size.
Keywords: Bragg peak degradation; dose calculation; lung tissue effect; proton therapy.
© 2020 Institute of Physics and Engineering in Medicine.
Similar articles
-
On the Way to Accounting for Lung Modulation Effects in Particle Therapy of Lung Cancer Patients-A Review.Cancers (Basel). 2024 Oct 25;16(21):3598. doi: 10.3390/cancers16213598. Cancers (Basel). 2024. PMID: 39518037 Free PMC article. Review.
-
Benchmarking a GATE/Geant4 Monte Carlo model for proton beams in magnetic fields.Med Phys. 2020 Jan;47(1):223-233. doi: 10.1002/mp.13883. Epub 2019 Nov 13. Med Phys. 2020. PMID: 31661559 Free PMC article.
-
A pencil beam algorithm for magnetic resonance image-guided proton therapy.Med Phys. 2018 May;45(5):2195-2204. doi: 10.1002/mp.12854. Epub 2018 Mar 30. Med Phys. 2018. PMID: 29532490 Free PMC article.
-
Evaluation of monte carlo to support commissioning of the treatment planning system of new pencil beam scanning proton therapy facilities.Phys Med Biol. 2024 Feb 13;69(4). doi: 10.1088/1361-6560/ad1272. Phys Med Biol. 2024. PMID: 38052092
-
The effect of voxelization in Monte Carlo simulation to validate Bragg peak characteristics for a pencil proton beam.Rep Pract Oncol Radiother. 2023 Apr 6;28(1):102-113. doi: 10.5603/RPOR.a2023.0007. eCollection 2023. Rep Pract Oncol Radiother. 2023. PMID: 37122904 Free PMC article. Review.
Cited by
-
Dose calculation accuracy in particle therapy: Comparing carbon ions with protons.Med Phys. 2021 Nov;48(11):7333-7345. doi: 10.1002/mp.15209. Epub 2021 Sep 23. Med Phys. 2021. PMID: 34482555 Free PMC article.
-
On the Way to Accounting for Lung Modulation Effects in Particle Therapy of Lung Cancer Patients-A Review.Cancers (Basel). 2024 Oct 25;16(21):3598. doi: 10.3390/cancers16213598. Cancers (Basel). 2024. PMID: 39518037 Free PMC article. Review.
-
A review of the clinical introduction of 4D particle therapy research concepts.Phys Imaging Radiat Oncol. 2024 Jan 10;29:100535. doi: 10.1016/j.phro.2024.100535. eCollection 2024 Jan. Phys Imaging Radiat Oncol. 2024. PMID: 38298885 Free PMC article. Review.
-
The Influence of Motion on the Delivery Accuracy When Comparing Actively Scanned Carbon Ions versus Protons at a Synchrotron-Based Radiotherapy Facility.Cancers (Basel). 2022 Mar 31;14(7):1788. doi: 10.3390/cancers14071788. Cancers (Basel). 2022. PMID: 35406558 Free PMC article.
MeSH terms
Substances
LinkOut - more resources
Full Text Sources