Topology optimization of a mandibular reconstruction plate and biomechanical validation
- PMID: 33187871
- DOI: 10.1016/j.jmbbm.2020.104157
Topology optimization of a mandibular reconstruction plate and biomechanical validation
Abstract
Objectives: Reconstruction plates, used to bridge segmental defects of the mandible after tumor resection or traumatic bone tissue loss, are subjected to repeated stresses of mastication. High stress concentrations in these plates can result in hardware failure. Topology optimization (TO) could reduce the peak stress by computing the most optimal material distribution in a patient-specific implant (PSI) used for mandibular reconstruction. The objective of this study was biomechanical validation of a TO-PSI.
Methods: A computer-aided design (CAD) model with a segmental defect was created based on the geometry of a polyurethane mandible model. A standard-PSI was designed to bridge the defect. A TO-PSI was then designed with a maximum stress equal to the ultimate tensile stress of Ti6Al4V (930 MPa) during a loading condition of 378 N. Finite element analysis (FEA) was used to analyze stresses in both PSI designs during loading. The standard-PSI and TO-PSI designs were produced in triplicate by selective laser melting of Ti6Al4V, fixated to polyurethane mandible models with segmental defects identical to the CAD model, and subsequently subjected to continuous compression with a speed of 1 mm/min on a universal testing machine, while recording the load. Peak loads before failure in the TO-PSI group within a 30% range of the predicted peak load (378 N) were considered a successful biomechanical validation.
Results: Fracture of the TO-PSI occurred at a median peak load of 334 N (range 304-336 N). These values are within the 30% range of the predicted peak load. Fracture of the mandible model in the standard-PSI group occurred at a median peak load of 1100 N (range 1010-1460 N). Failure locations during biomechanical testing of TO-PSI and standard-PSI samples corresponded to regions in the FEA where stresses exceeded the ultimate tensile strength of titanium and polyurethane, respectively.
Conclusion: This study demonstrates a successful preliminary biomechanical validation of TO in the design process for mandibular reconstruction plates. Further work is needed to refine the finite element model, which is necessary to ultimately design TO-PSIs for clinical use.
Keywords: Biomechanical validation; Finite element analysis; Mandibular reconstruction; Patient-specific implant; Topology optimization.
Copyright © 2020 The Authors. Published by Elsevier Ltd.. All rights reserved.
Similar articles
-
Biomechanical validation of structural optimized patient-specific mandibular reconstruction plate orienting additive manufacturing.Comput Methods Programs Biomed. 2022 Sep;224:107023. doi: 10.1016/j.cmpb.2022.107023. Epub 2022 Jul 14. Comput Methods Programs Biomed. 2022. PMID: 35872386
-
Biomechanical evaluation of a novel hybrid reconstruction plate for mandible segmental defects: A finite element analysis and fatigue testing.J Craniomaxillofac Surg. 2017 Oct;45(10):1671-1680. doi: 10.1016/j.jcms.2017.07.010. Epub 2017 Jul 27. J Craniomaxillofac Surg. 2017. PMID: 28838835
-
The biomechanical aspects of reconstruction for segmental defects of the mandible: a finite element study to assess the optimisation of plate and screw factors.J Craniomaxillofac Surg. 2014 Sep;42(6):855-62. doi: 10.1016/j.jcms.2013.12.005. Epub 2013 Dec 31. J Craniomaxillofac Surg. 2014. PMID: 24467871
-
Computational models and their applications in biomechanical analysis of mandibular reconstruction surgery.Comput Biol Med. 2024 Feb;169:107887. doi: 10.1016/j.compbiomed.2023.107887. Epub 2023 Dec 21. Comput Biol Med. 2024. PMID: 38160502 Review.
-
Patient-specific finite element models of the human mandible: Lack of consensus on current set-ups.Oral Dis. 2021 Jan;27(1):42-51. doi: 10.1111/odi.13381. Epub 2020 Jul 9. Oral Dis. 2021. PMID: 32372548 Free PMC article. Review.
Cited by
-
Mandibular Implants: A Metamaterial-Based Approach to Reducing Stress Shielding.Adv Healthc Mater. 2025 May;14(13):e2500405. doi: 10.1002/adhm.202500405. Epub 2025 Apr 4. Adv Healthc Mater. 2025. PMID: 40183751 Free PMC article.
-
Measurement of Thickness at the Inferior Border of the Mandible Using Computed Tomography Images: A Retrospective Study including 300 Japanese Cases.Tomography. 2023 Jun 22;9(4):1236-1245. doi: 10.3390/tomography9040098. Tomography. 2023. PMID: 37489466 Free PMC article.
-
Biomechanical assessment of mandibular fracture fixation using finite element analysis validated by polymeric mandible mechanical testing.Sci Rep. 2024 May 23;14(1):11795. doi: 10.1038/s41598-024-62011-4. Sci Rep. 2024. PMID: 38782942 Free PMC article.
-
Finite element analysis of mandibular fracture fixation authenticated by 3D printed mandible mechanical testing.Sci Rep. 2025 Apr 26;15(1):14655. doi: 10.1038/s41598-025-98732-3. Sci Rep. 2025. PMID: 40287549 Free PMC article.
-
Topology Optimization: A Review for Structural Designs Under Statics Problems.Materials (Basel). 2024 Dec 6;17(23):5970. doi: 10.3390/ma17235970. Materials (Basel). 2024. PMID: 39685406 Free PMC article. Review.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Miscellaneous