. 2021 Mar;44(1):23-35.
doi: 10.1007/s13246-020-00948-2. Epub 2020 Nov 23.

Technical evaluation of image quality in synthetic mammograms obtained from 15° and 40° digital breast tomosynthesis in a commercial system: a quantitative comparison

Affiliations

Technical evaluation of image quality in synthetic mammograms obtained from 15° and 40° digital breast tomosynthesis in a commercial system: a quantitative comparison

Patrizio Barca et al. Phys Eng Sci Med. 2021 Mar.

Abstract

Digital breast tomosynthesis (DBT) has recently gained interest both for breast cancer screening and diagnosis. Its employment has increased also in conjunction with digital mammography (DM), to improve cancer detection and reduce false positive recall rate. Synthetic mammograms (SMs) reconstructed from DBT data have been introduced to replace DM in the DBT + DM approach, for preserving the benefits of the dual-acquisition modality whilst reducing radiation dose and compression time. Therefore, different DBT models have been commercialized and the effective potential of each system has been investigated. In particular, wide-angle DBT was shown to provide better depth resolution than narrow-angle DBT, while narrow-angle DBT allows better identification of microcalcifications compared to wide-angle DBT. Given the increasing employment of SMs as supplement to DBT, a comparison of image quality between SMs obtained in narrow-angle and wide-angle DBT is of practical interest. Therefore, the aim of this phantom study was to evaluate and compare the image quality of SMs reconstructed from 15° (SM15) and 40° (SM40) DBT in a commercial system. Spatial resolution, noise and contrast properties were evaluated through the modulation transfer function (MTF), noise power spectrum, maps of signal-to-noise ratio (SNR), image contrast, contrast-to-noise ratio (CNR) and contrast-detail (CD) thresholds. SM40 expressed higher MTF than SM15, but also lower SNR and CNR levels. SM15 and SM40 were characterized by slight different texture, and a different behavior in terms of contrast was found. SM15 provided better CD performances than SM40. These results suggest that the employment of wide/narrow-angle DBT + SM images should be optimized based on the specific image task.

Keywords: Digital breast tomosynthesis; Image quality; Mammography; S-view; Synthetic mammogram; Wide angle DBT.

PubMed Disclaimer

References

    1. Lauby-Secretan B, Scoccianti C, Loomis D et al (2015) Breast-cancer screening: viewpoint of the IARC Working Group. N Engl J Med 372(24):2353–2358 - PubMed
    1. Hooley RJ, Durand MA, Philpotts LE (2017) Advances in digital breast tomosynthesis. AJR Am J Roentgenol 208(2):256–266 - PubMed
    1. Pattacini P, Nitrosi A, Giorgi Rossi P et al (2018) Digital mammography versus digital mammography plus tomosynthesis for breast cancer screening: the Reggio Emilia tomosynthesis randomized trial. Radiology 288(2):375–385 - PubMed
    1. European Reference Organisation for Quality Assured Breast Screening and Diagnostic Services (2018) Protocol for the quality control of the physical and technical aspects of digital breast tomosynthesis systems. (EUREF)
    1. Badano A, Graff CG, Badal A et al (2018) Evaluation of digital breast tomosynthesis as replacement of full-field digital mammography using an in silico imaging trial. JAMA Netw Open 1(7):e185474 - PubMed - PMC

LinkOut - more resources