Comparison of the SARS-CoV-2 Rapid antigen test to the real star Sars-CoV-2 RT PCR kit
- PMID: 33227341
- PMCID: PMC7678421
- DOI: 10.1016/j.jviromet.2020.114024
Comparison of the SARS-CoV-2 Rapid antigen test to the real star Sars-CoV-2 RT PCR kit
Abstract
There is an ongoing need for reliable antigen assays for timely and easy detection of individuals with acute SARS-CoV-2 infection. Using 75 swabs from patients previously tested positive by SARS-CoV-2 PCR and 75 swabs from patients previously tested negative by SARS-CoV-2 PCR, we investigated the sensitivity and specificity of the SARS-CoV-2 Rapid Antigen Test (Roche). We determined a specificity of 96 %. The assay's sensitivity with samples with a cycle threshold of < 25, 25 - <30, 30 - <35, and> = 35 was 100 %, 95 %, 44.8 % and 22.2 %, respectively. We conclude that sensitivity and specificity of the antigen assay is inferior to the PCR assay. However, the antigen assay may be a quick and easy to perform alternative for differentiation of individuals contagious for SARS-CoV-2 from non-contagious individuals.
Keywords: Antigen; COVID-19; Lateral flow assay; SARS-CoV-2.
Copyright © 2020 Elsevier B.V. All rights reserved.
Figures
References
-
- Eis-Hübinger A.M., Hönemann M., Wenzel J.J., Berger A., Widera M., Schmidt B., Aldabbagh S., Marx B., Streeck H., Ciesek S., Liebert U.G., Huzly D., Hengel H., Panning M. Ad hoc laboratory-based surveillance of SARS-CoV-2 by real-time RT-PCR using minipools of RNA prepared from routine respiratory samples. J. Clin. Virol. 2020;127:104381. - PMC - PubMed