Repeatability of Orbscan III for anterior segment parameters in normal eyes
- PMID: 33229667
- PMCID: PMC7856959
- DOI: 10.4103/ijo.IJO_2856_20
Repeatability of Orbscan III for anterior segment parameters in normal eyes
Abstract
Purpose: Assessment of repeatability of Orbscan III in measuring steep keratometry (sKm), flat keratometry (fKm), thinnest corneal thickness (TCT), anterior chamber depth (ACD), white-to-white diameter (WTW) and 3 mm and 5 mm zonal irregularity in healthy eyes.
Methods: In this prospective observatioanl study 100 eyes of 50 participants underwent three consecutive scans on Orbscan III by a single examiner with a gap of 5 minutes between scans. Statistical analysis was performed to assess the repeatability of sKm, fKm, TCT, WTW, ACD and parameters of 3- and 5 mm irregularity. Within subject standard deviation (Sw), coefficient of variation (CoV) and test retest variability (TRT) was calculated for the commonly used parameters.
Were the main outcome measures.: The repeatability of sKm, fKm, TCT, WTW, ACD, 3 mm and 5 mm zonal irregularity on Orbscan III.
Results: Total of 50 patients (25 male and 25 female) who fulfilled inclusion and exclusion criteria were assessed. sKm, fKm, TCT and WTW were found to have a TRT of 0.31, 0.29. 21.5 and 0.27, respectively, and the ICC values were found to be over 0.9, denoting high repeatability. Repeat measurements of parameters of irregularity and ACD were found to be statistically similar, again denoting similarity between repeat measurements.
Conclusion: Orbscan III provides repeatable measurements of Sim-k, anterior chamber depth, thinnest corneal thickness, 3 mm and 5 mm zonal irregularity and white-to-white diameter in healthy eyes.
Keywords: Corneal parameters; Orbscan III; repeatability.
Conflict of interest statement
None
Comment in
-
Comments on: Repeatability of Orbscan III for anterior segment parameters in normal eyes.Indian J Ophthalmol. 2021 May;69(5):1341-1342. doi: 10.4103/ijo.IJO_144_21. Indian J Ophthalmol. 2021. PMID: 33913902 Free PMC article. No abstract available.
-
Response to comments on: Repeatability of Orbscan III for anterior segment parameters in normal eyes.Indian J Ophthalmol. 2021 May;69(5):1342-1343. doi: 10.4103/ijo.IJO_490_21. Indian J Ophthalmol. 2021. PMID: 33913903 Free PMC article. No abstract available.
References
-
- Mashige KP. A review of corneal diameter, curvature and thickness values and influencing factors. African Vision and Eye Health. 2013;72:185–94.
-
- Martin R, Izquierdo M, De Juan V, Rodriguez G, Fernandezm I. Repeatability and reproducibility of Orbscan II. Optometry Rep. 2012:2. doi: 104081/optometry 2012e1.
-
- Eyewire. Bausch + Lomb Introduces Orbscan 3 Anterior Segment Analyzer. 2016. [Last accessed on 2020 Sep 02]. Available from: https://eyewirenews/articles/bausch-lomb-introduces-orbscan-3-anterior-s...
-
- Wei RH, Lim L, Chan WK, Tan DT. Evaluation of Orbscan II corneal topography in individuals with myopia. Ophthalmology. 2006;113:177–83. - PubMed
MeSH terms
LinkOut - more resources
Full Text Sources
Miscellaneous