. 2021 Jul;41(4):427-431.
doi: 10.1177/0896860820973120. Epub 2020 Nov 28.

Intraperitoneal pressure: Stability over time and validation of Durand's measurement method

Affiliations

Intraperitoneal pressure: Stability over time and validation of Durand's measurement method

Alicia Sobrino-Pérez et al. Perit Dial Int. 2021 Jul.

Abstract

Intraperitoneal pressure (IPP) is gaining consideration as a relevant parameter of peritoneal dialysis (PD) in adults, although many of its aspects are still pending clarification. We address here its stability over time and the validity of the usual method of clinical measurement, as proposed by Durand in 1992 but never specifically validated. We performed this validation by comparing Durand's method and direct measurements with a central venous pressure system. We performed a total of 250 measurement pairs in 50 patients with different intraperitoneal volumes plus in-vitro measurements with a simulated peritoneum. Absolute differences between the two systems in vivo were 0.87 ± 0.91 cmH2O (range 0-5 cmH2O); only 6.4% of them were ≥3 cmH2O. In vitro results for both methods were identical. We also compared IPP measurements in the same patient separated by 1-4 h (514 measurement pairs in 136 patients), 1 week (92 pairs in 92 patients), and 2 years (34 pairs in 17 patients). Net differences of measurements separated by hours or 1 week were close to 0 cmH2O, with oscillations of 1.5 cmH2O in hours and 2.3 cmH2O in 1 week. IPP measured 2 years apart presented a net decrease of 2.5 ± 4.9 cmH2O, without correlation with body mass index changes or any other usual parameter of PD. In hours, 7% of IPP differences were >3 cmH2O, 22% in 1 week, and 50% in 2 years. In conclusion, Durand's method is precise enough to measure IPP in peritoneal dialysis. This parameter is not stable over long timescales, so it is necessary to use recent measurements.

Keywords: Durand’s method; intraperitoneal pressure; peritoneal dialysis; stability.