Does the minimal occlusive volume technique provide adequate endotracheal tube cuff pressure to prevent air leakage?: a prospective, randomized, crossover clinical study
- PMID: 33329837
- PMCID: PMC7713835
- DOI: 10.17085/apm.20020
Does the minimal occlusive volume technique provide adequate endotracheal tube cuff pressure to prevent air leakage?: a prospective, randomized, crossover clinical study
Abstract
Background: Methods of determining proper endotracheal tube (ETT) cuff pressure to prevent air leakage include the minimal occlusive volume (MOV) technique, which uses auscultation, and the spirometer technique, which directly measures inspiratory and expiratory breathing volumes. Spirometers may measure even small air leakage, therefore, the spirometer technique requires a higher cuff pressure than the MOV technique to completely seal the airway. This study aimed to evaluate the difference in cuff pressure between the two techniques used to seal the airway.
Methods: Thirty-five female patients were intubated using an ETT with a cuff, and cuff inflation was performed with both techniques at a 10-min interval in random order-the MOV technique and then the spirometer technique or vice versa. The cuff pressure was measured at each period.
Results: The cuff pressures were 16.7 ± 4.4 cmH2O and 18.7 ± 5.2 cmH2O for the MOV and spirometer techniques, respectively. The cuff pressure for the spirometer technique was 2.0 cmH2O higher than that for the MOV technique and this difference was statistically significant (95% confidence interval, 0.7-3.3; P = 0.003). Considering the upper end (3.3 cmH2O) of the 95% confidence interval and the size of one scale unit (2.0 cmH2O) of a manometer, the difference in cuff pressure was up to 4 cmH2O in practice.
Conclusions: Even though the air leakage sound disappears on auscultation, unlike the previous recommendation, the airway sealing would be completed only by increasing the cuff pressure by approximately 4 cmH2O.
Keywords: Airway management; Anesthesia; Endotracheal; Manometry; Minimal occlusive volume; Spirometer.
Copyright © the Korean Society of Anesthesiologists, 2020.
Conflict of interest statement
CONFLICTS OF INTEREST No potential conflict of interest relevant to this article was reported.
Figures
Similar articles
-
Risk factor evaluation of cuff pressure of >30 cmH2O to stop air leakage during mechanical ventilation: A prospective observational study.Nurs Open. 2024 Jun;11(6):e2187. doi: 10.1002/nop2.2187. Nurs Open. 2024. PMID: 38837558 Free PMC article.
-
The evaluation of endotracheal tube cuff pressure in pediatric patients by subjective inflation techniques: a prospective observational study.BMC Anesthesiol. 2024 Nov 5;24(1):399. doi: 10.1186/s12871-024-02780-7. BMC Anesthesiol. 2024. PMID: 39501142 Free PMC article.
-
Comparison of four inflation techniques on endotracheal tube cuff pressure using a feline airway simulator.J Feline Med Surg. 2020 Jul;22(7):641-647. doi: 10.1177/1098612X19871701. Epub 2019 Aug 29. J Feline Med Surg. 2020. PMID: 31464541 Free PMC article.
-
Passive Release Technique Produces the Most Accurate Endotracheal Tube Cuff Pressure Than Manual Palpation and Minimum Occlusive Volume Technique in the Absence of Manometer.Turk J Anaesthesiol Reanim. 2021 Apr;49(2):114-117. doi: 10.5152/TJAR.2020.80. Epub 2020 Nov 30. Turk J Anaesthesiol Reanim. 2021. PMID: 33997839 Free PMC article.
-
Aspects of in vivo endotracheal tube intracuff pressure in cats.Vet Anaesth Analg. 2019 Jan;46(1):55-63. doi: 10.1016/j.vaa.2018.06.017. Epub 2018 Oct 13. Vet Anaesth Analg. 2019. PMID: 30553724
Cited by
-
Risk factor evaluation of cuff pressure of >30 cmH2O to stop air leakage during mechanical ventilation: A prospective observational study.Nurs Open. 2024 Jun;11(6):e2187. doi: 10.1002/nop2.2187. Nurs Open. 2024. PMID: 38837558 Free PMC article.
-
Relationship between difference in endotracheal tube cuff area and airway area with minimum cuff pressure for adequate airway sealing: a prospective observational study.Sci Rep. 2025 Feb 18;15(1):5875. doi: 10.1038/s41598-025-85355-x. Sci Rep. 2025. PMID: 39966434 Free PMC article.
-
Effect of Minimum Bronchial Cuff Volume of Left-Sided Double-Lumen Tube for One-Lung Ventilation on the Change in Bronchial Cuff Pressure during Lateral Positioning in Thoracic Surgery: A Prospective Observational Study.J Clin Med. 2023 Mar 24;12(7):2473. doi: 10.3390/jcm12072473. J Clin Med. 2023. PMID: 37048557 Free PMC article.
References
-
- Vyas D, Inweregbu K, Pittard A. Measurement of tracheal tube cuff pressure in critical care. Anaesthesia. 2002;57:275–7. - PubMed
-
- Mahul P, Auboyer C, Jospe R, Ros A, Guerin C, el Khouri Z, et al. Prevention of nosocomial pneumonia in intubated patients: respective role of mechanical subglottic secretions drainage and stress ulcer prophylaxis. Intensive Care Med. 1992;18:20–5. - PubMed
-
- Mandøe H, Nikolajsen L, Lintrup U, Jepsen D, Mølgaard J. Sore throat after endotracheal intubation. Anesth Analg. 1992;74:897–900. - PubMed
-
- Curiel García JA, Guerrero-Romero F, Rodríguez-Morán M. [Cuff pressure in endotracheal intubation: should it be routinely measured?] Gac Med Mex. 2001;137:179–82. Spanish. - PubMed
-
- Hähnel J, Treiber H, Konrad F, Eifert B, Hahn R, Maier B, et al. [A comparison of different endotracheal tubes. Tracheal cuff seal, peak centering and the incidence of postoperative sore throat] Anaesthesist. 1993;42:232–7. German. - PubMed
LinkOut - more resources
Full Text Sources