Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
. 2020 Nov 30:7:587945.
doi: 10.3389/fcvm.2020.587945. eCollection 2020.

Cardiac Ambulatory Monitoring: New Wireless Device Validated Against Conventional Holter Monitoring in a Case Series

Affiliations

Cardiac Ambulatory Monitoring: New Wireless Device Validated Against Conventional Holter Monitoring in a Case Series

Srinivasan Murali et al. Front Cardiovasc Med. .

Abstract

Background: Cardiac arrhythmias are very common but underdiagnosed due to their transient and asymptomatic nature. An optimization of arrhythmia detection would permit to better treat patients and could substantially reduce morbidity and mortality. The SmartCardia ScaAI wireless patch is a novel CE IIa approved, single-lead electrocardiographic (ECG) ambulatory monitor designed for cardiac arrhythmias detection. Hypothesis: The accuracy of the new SmartCardia wireless patch to detect arrhythmias is comparable to the conventional Holter monitoring. Methods: Patients referred for a suspicion of arrhythmia between February and March 2020 were included in the trial. Simultaneous ambulatory ECG were recorded using a conventional 24-h Holter and the SmartCardia. The primary endpoint was the detection of cardiac arrhythmias over the total wear time of the devices, defined as premature atrial contraction (PAC), supraventricular tachycardia ≥3 beats, premature ventricular contraction (PVC), and ventricular tachycardia ≥3 beats. Conduction abnormalities, pause ≥2 s and atrioventricular block (AVB), were also tracked. McNemar's test was used to compare the matched pairs of data from both devices. Results: A total of 40 patients were included in the trial. Over the total wear time, there was no significant difference between the devices for ventricular and supraventricular arrhythmias detection. Pauses and AVB were equally identified by the two devices in three patients. Conclusion: Over the total wear time, the SmartCardia device showed an accuracy to detect arrhythmia similar to the 24-h Holter monitoring: single-lead, adhesive-patch monitoring might become an interesting alternative to the conventional Holter monitoring.

Keywords: Holter monitoring; SmartCardia patch; atrial fibrillation; cardiac arrhythmias; electrocardiogram; wireless recorder.

PubMed Disclaimer

Conflict of interest statement

SM, FR, and MM are members of the board of SmartCardia. The remaining authors declare that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.

Figures

Figure 1
Figure 1
(A) Smartcardia Wireless Disposable, (B) PatchDevice patched on the patient's chest.
Figure 2
Figure 2
Comparison of the average heart rate between the two systems showing good correlation.
Figure 3
Figure 3
Bland-Altman plots for agreement between heart rate evaluation by the SmartCardia and 24-h Holter monitor. (A) Minimal, (B) maximal, and (C) average heart rate. HR, Heart Rate; BPM, Beat Per Minute; min, minimal; avrg, average; max, maximal.
Figure 4
Figure 4
Examples of traces from both devices. (A) PAC with some aberrant conduction correctly classified by the SmartCardia (the Holter monitoring classified the premature beats as PVC), (B) Short VT correctly classified by the SmartCardia, (C) Recording of the same episode of supraventricular tachycardia (above Holter, below SmartCardia).

Similar articles

Cited by

References

    1. Hart RG. Stroke prevention in atrial fibrillation. Curr Cardiol Rep. (2000) 2:51–5. 10.1007/s11886-000-0025-2 - DOI - PubMed
    1. Jabaudon D, Sztajzel J, Sievert K, Landis T, Sztajzel R. Usefulness of ambulatory 7-day ECG monitoring for the detection of atrial fibrillation and flutter after acute stroke and transient ischemic attack. Stroke. (2004) 35:1647–51. 10.1161/01.STR.0000131269.69502.d9 - DOI - PubMed
    1. Joshi AK, Kowey PR, Prystowsky EN, Benditt DG, Cannom DS, Pratt CM, et al. . First experience with a Mobile Cardiac Outpatient Telemetry (MCOT) system for the diagnosis and management of cardiac arrhythmia. Am J Cardiol. (2005) 95:878–81. 10.1016/j.amjcard.2004.12.015 - DOI - PubMed
    1. Olson JA, Fouts AM, Padanilam BJ, Prystowsky EN. Utility of mobile cardiac outpatient telemetry for the diagnosis of palpitations, presyncope, syncope, and the assessment of therapy efficacy. J Cardiovasc Electrophysiol. (2007) 18:473–7. 10.1111/j.1540-8167.2007.00779.x - DOI - PubMed
    1. Ritter MA, Kochhauser S, Duning T, Reinke F, Pott C, Dechering DG, et al. Occult atrial fibrillation in cryptogenic stroke: detection by 7-day electrocardiogram versus implantable cardiac monitors. Stroke. (2013) 44:1449–52. 10.1161/STROKEAHA.111.676189 - DOI - PubMed

LinkOut - more resources