Effect of Body Position on Intraocular Pressure Measured by Rebound Tonometer in Healthy Children

Affiliations

Effect of Body Position on Intraocular Pressure Measured by Rebound Tonometer in Healthy Children

Dilek Uzlu et al. Turk J Ophthalmol. .

Abstract

Objectives: To evaluate the effect of body position on intraocular pressure (IOP) measurement in the pediatric age group.

Materials and methods: Children whose general condition was healthy and ophthalmic examination was within normal limits were included. Forty-nine eyes of 49 pediatric patients were included in the study. IOP was measured with an ICARE rebound tonometer (ICARE PRO; ICARE, Helsinki, Finland) while patients were in standing, sitting, and supine positions. Differences between the consecutive measurements were compared statistically.

Results: Twenty-two of the 49 patients were female, 27 were male. The mean age was 9.61±2.66 (5-15) years. Mean IOP values in the standing, sitting, and supine positions were 18.81±2.97 (11.6-26.2) mmHg, 18.88±3.44, (12-28.2) mmHg, and 19.01±2.8 (13.5-25.9) mmHg, respectively. There were no statistically significant differences in pairwise comparisons of the measurements taken in the different positions (p=0.846, p=0.751, p=0.606). There was a statistically significant correlation between corneal thickness and intraocular pressure values in all measurements (p=0.001, r=0.516).

Conclusion: IOP values measured with the ICARE rebound tonometer in healthy children are not affected by body position.

Keywords: Child; body position; intraocular pressure; ocular tonometry.

PubMed Disclaimer

Conflict of interest statement

Conflict of Interest: No conflict of interest was declared by the authors.

References

    1. Heijl A, Leske MC, Bengtsson B, Hyman L, Bengtsson B, Hussein M; Early Manifest Glaucoma Trial Group. Reduction of intraocular pressure and glaucoma progression: results from the Early Manifest Glaucoma Trial. Arch Ophthalmol. 2002;120:1268–1279. - PubMed
    1. Leske MC, Heijl A, Hussein M, Bengtsson B, Hyman L, Komaroff E; Early Manifest Glaucoma Trial Group. Factors for glaucoma progression and the effect of treatment: the early manifest glaucoma trial. Arch Ophthalmol. 2003;121:48–56. - PubMed
    1. Pakrou N, Gray T, Mills R, Landers J, Craig J. Clinical comparison of the ICare tonometer and Goldmann applanation tonometry. J Glaucoma. 2008;17:43–47. - PubMed
    1. Brusini P, Salvetat ML, Zeppieri M, Tosoni C, Parisi L. Comparison of ICare tonometer with Goldmann applanation tonometer in glaucoma patients. J Glaucoma. 2006;15:213–217. - PubMed
    1. Iliev ME, Goldblum D, Katsoulis K, Amstutz C, Frueh B. Comparison of rebound tonometry with Goldmann applanation tonometry and correlation with central corneal thickness. Br J Ophthalmol. 2006;90:833–835. - PMC - PubMed