. 2020 Dec;6(14):1744-1751.
doi: 10.1016/j.jacep.2020.06.010. Epub 2020 Aug 26.

Retrieval of Long-Term Implanted Leadless Pacemakers: A Single-Center Experience

Affiliations
Free article

Retrieval of Long-Term Implanted Leadless Pacemakers: A Single-Center Experience

Kentaro Minami et al. JACC Clin Electrophysiol. 2020 Dec.
Free article

Abstract

Objectives: The authors report their single-center experience with the retrieval and replacement of long-term implanted leadless cardiac pacemakers (LPs) and the factors contributing to success.

Background: LPs are a clinically effective and safe alternative to standard transvenous pacemakers for single-chamber ventricular pacing. However, the feasibility of retrieving and replacing long-term implanted LPs is not well known.

Methods: A total of 34 patients with implanted Nanostim LPs (mean implantation duration 1,570 ± 479 days) subsequently underwent retrieval. On the basis of fluoroscopic imaging, the cohort was divided into 2 groups: those with remarkable swinging movement (SM) of the LP docking button (n = 25) and those without SM (n = 9).

Results: The overall LP retrieval success rate was 85% (n = 29). New leadless devices were implanted in 27 patients immediately after LP retrieval. No procedure-related adverse events occurred. The rate of successful LP retrieval was significantly higher in the SM group (25 of 25 [100%]) than in the no-SM group (4 of 9 [44%]) (p < 0.001), and fluoroscopy time during the retrieval procedure was shorter in the SM group (12.7 ± 8.6 min) than in the no-SM group (45.6 ± 19 min) (p < 0.001).

Conclusions: This study demonstrated the feasibility and safety of retrieval of long-term implanted LPs after a mean duration of 4 years. Retrieval was most successful in patients whose docking buttons exhibited significant SM.

Keywords: Nanostim; leadless cardiac pacemaker; pacemaker retrieval.

PubMed Disclaimer

Conflict of interest statement

Author Disclosures This study was supported by the Ministry of Health, Czech Republic, DRO (NHH, 00023884). Drs. Neužil and Reddy have received grant support from Abbott (St. Jude Medical), the manufacturer of the Nanostim leadless pacemaker. All other authors have reported that they have no relationships relevant to the contents of this paper to disclose.