Initial Experience, Safety, and Feasibility of Left Bundle Branch Area Pacing: A Multicenter Prospective Study

Affiliations
Free article
Multicenter Study

Initial Experience, Safety, and Feasibility of Left Bundle Branch Area Pacing: A Multicenter Prospective Study

Santosh K Padala et al. JACC Clin Electrophysiol. 2020 Dec.
Free article

Abstract

Objectives: This study sought to evaluate the safety and feasibility of conduction system pacing by performing left bundle branch area pacing (LBBAP).

Background: There are limited data from single centers showing that LBBAP may circumvent the technical and electrophysiological challenges encountered with His bundle pacing.

Methods: Patients referred for pacemaker implantation at 2 centers between February 1, 2019, and March 31, 2020, were considered for LBBAP. LBBAP was performed by implanting a lumen-less, exposed helix lead approximately 2 cm distal to the His bundle and deep into the septum using a specialized delivery sheath. Implant success rates, complications, and electrophysiological parameters were assessed.

Results: LBBAP was successful in 305 of 341 patients (89%). Mean age was 72 ± 12 years; 45% were women; and 39% had QRS duration (QRSd) >130 ms, 22% right bundle branch block, 11% left bundle branch block, and 6% intraventricular conduction defect. Pacing indications were sinus node dysfunction in 28.7%, atrioventricular block in 52.5%, cardiac resynchronization therapy in 8.8%, and refractory atrial fibrillation in 10% of patients. Procedural duration was 74.7 ± 34 min and fluoroscopic time was 10.4 ± 8.1 min. The mean baseline QRSd and paced QRSd in the overall cohort was 114 ± 29.8 ms versus 112 ± 11.7 ms (p < 0.001) and in patients with infra-Hisian disease was 144.5 ± 19 ms versus 115 ± 12 ms (p < 0.001), respectively. Mean left ventricular activation time was 71.7 ± 11 ms at high output and 74.7 ± 11 ms at low output. LBB potentials were noted in 41% patients. Pacing threshold and R waves were 0.74 ± 0.3 V at 0.4 ms and 10.7 ± 4.9 mV at time of implantation and were stable at 1-, 3-, 6-, and 12-month follow-ups. The only major complications were 3 LBBAP lead dislodgements, 2 within 24 h and 1 at 2 weeks.

Conclusions: LBBA pacing is safe, feasible, and a reliable alternative to His bundle pacing for providing physiological pacing. Randomized controlled studies are needed to confirm the safety, feasibility, and clinical outcomes of LBBAP.

Keywords: His bundle pacing; conduction system pacing; left bundle branch area pacing; physiological pacing.

PubMed Disclaimer

Conflict of interest statement

Author Disclosures Dr. Padala has served as a consultant for Medtronic. Dr. Koneru has received teaching honoraria from Medtronic and Biotronik; and has received fellowship support from Boston Scientific, Biosense Webster, Medtronic, and Abbott Medical. Dr. Ellenbogen has served as a consultant and Data and Safety Monitoring Board chair for Medtronic; and honoraria and research support from Medtronic. Dr. Verma has served as a consultant for Medtronic, Biosense Webster, Boston Scientific, and Servier; and has received grant support from Bayer and Biotronik. All other authors have reported that they have no relationships relevant to the contents of this paper to disclose.