Precision error in dual-photon absorptiometry related to source age
- PMID: 3336729
- DOI: 10.1148/radiology.166.2.3336729
Precision error in dual-photon absorptiometry related to source age
Abstract
An average, variable precision error of up to 6% related to source age was observed for dual-photon absorptiometry of the spine in a longitudinal study of bone mineral content involving 393 women. Application of a software correction for source decay compensated for only a portion of this error. The authors conclude that measurement of bone-loss rates using serial dual-photon bone mineral measurements must be interpreted with caution.