Wear and migration are not influenced by head size in a vitamin E-infused highly cross-linked polyethylene acetabular cup
- PMID: 33384276
- DOI: 10.1016/j.otsr.2020.03.030
Wear and migration are not influenced by head size in a vitamin E-infused highly cross-linked polyethylene acetabular cup
Abstract
Background: Aseptic loosening and periprosthetic osteolysis are frequent complications in total hip arthroplasty requiring revision surgery. Highly cross-linked polyethylene (HXLPE) implants have improved wear resistance, permitting larger femoral heads. However, such implants may experience surface cracking, mechanical failure, and oxidative damage. Vitamin E-infused HXLPE (VEPE) implants were therefore developed to reduce oxidation without compromising mechanical strength. We addressed the following questions: (1) Does femoral head size affect the midterm annual polyethylene wear rates of VEPE acetabular cups? (2) Does femoral head size affect the midterm migration rates of VEPE acetabular cups? (3) Are clinical outcomes affected by femoral head size?
Hypothesis: Annual wear rate, migration rate, and clinical outcomes of VEPE acetabular cups are independent of femoral head size.
Patients and methods: This was a prospective, multicentre, observational study of patients that underwent total hip arthroplasty. Hips were grouped according to the size of femoral head implanted (28 mm, 32 mm, and 36 mm). We determined annual wear rate and migration rate of VEPE acetabular cups using the Einzel-Bild-Röntgen-Analyse software. Clinically, we assessed the Harris Hip Score and visual analog score for pain and satisfaction.
Results: We followed 253 patients (267 hips) for a mean of 55.0±20.6 months in the 28 mm, 46.2±21.4 months in the 32 mm, and 43.8±22.6 months in the 36 mm group. The annual wear rate was 0.025 mm per year from 1 year to the last follow-up and remained similar between the groups (p>0.05). Also, mean two-dimensional migration rates did not exceed 0.05 mm from 2 years to the last follow-up and remained similar between the groups (p=0.355). Finally, clinical outcomes also did not differ between the groups (p>0.05). Two patients required revision surgery.
Discussion: Femoral head size did not influence midterm annual wear rate, migration rate, and clinical outcomes of VEPE acetabular cups. Furthermore, wear and cup migration rates were below the reported values leading to osteolysis and aseptic loosening. Nevertheless, studies with extended follow-up periods will be necessary to confirm these results in the long term.
Level of evidence: IV.
Keywords: EBRA; Migration; Pressfit cup; VEPE; Wear.
Copyright © 2020 The Authors. Published by Elsevier Masson SAS.. All rights reserved.
Similar articles
-
An isoelastic monoblock cup versus a modular metal-back cup: a matched-pair analysis of clinical and radiological results using Einzel-Bild-Röntgen-Analyse software.Arch Orthop Trauma Surg. 2024 Jan;144(1):493-500. doi: 10.1007/s00402-023-05058-8. Epub 2023 Sep 23. Arch Orthop Trauma Surg. 2024. PMID: 37740060 Free PMC article.
-
The concept of a cementless isoelastic monoblock cup made of highly cross-linked polyethylene infused with vitamin E: radiological analyses of migration and wear using EBRA and clinical outcomes at mid-term follow-up.BMC Musculoskelet Disord. 2021 Jan 23;22(1):107. doi: 10.1186/s12891-021-03981-8. BMC Musculoskelet Disord. 2021. PMID: 33485345 Free PMC article.
-
Results of a randomized controlled trial with five-year radiostereometric analysis results of vitamin E-infused highly crosslinked versus moderately crosslinked polyethylene in reverse total hip arthroplasty.Bone Joint J. 2020 Dec;102-B(12):1646-1653. doi: 10.1302/0301-620X.102B12.BJJ-2020-0721.R1. Bone Joint J. 2020. PMID: 33249906 Clinical Trial.
-
In Vivo Performance of Vitamin E Stabilized Polyethylene Implants for Total Hip Arthroplasty: A Review.J Arthroplasty. 2023 May;38(5):970-979. doi: 10.1016/j.arth.2022.11.010. Epub 2022 Dec 5. J Arthroplasty. 2023. PMID: 36481286 Review.
-
Vitamin E highly cross-linked polyethylene reduces mid-term wear in primary total hip replacement: a meta-analysis and systematic review of randomized clinical trials using radiostereometric analysis.EFORT Open Rev. 2021 Sep 14;6(9):759-770. doi: 10.1302/2058-5241.6.200072. eCollection 2021 Sep. EFORT Open Rev. 2021. PMID: 34667647 Free PMC article. Review.
Cited by
-
An isoelastic monoblock cup versus a modular metal-back cup: a matched-pair analysis of clinical and radiological results using Einzel-Bild-Röntgen-Analyse software.Arch Orthop Trauma Surg. 2024 Jan;144(1):493-500. doi: 10.1007/s00402-023-05058-8. Epub 2023 Sep 23. Arch Orthop Trauma Surg. 2024. PMID: 37740060 Free PMC article.
-
Ten-year clinical and radiological outcomes with a vitamin E-infused highly cross-linked polyethylene acetabular cup.Bone Jt Open. 2024 Oct 3;5(10):825-831. doi: 10.1302/2633-1462.510.BJO-2023-0179.R1. Bone Jt Open. 2024. PMID: 39357881 Free PMC article.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical
Research Materials