Safety of lung stereotactic ablative radiotherapy for the functioning of cardiac implantable electronic devices
Abstract
Background and purpose: The prevalence of patients with a cardiac implantable device (CIED) developing cancer and requiring a course of radiotherapy (RT) is increasing remarkably. Previously published reports agree that standard and conventionally fractionated RT is usually safe for CIEDs, but no "in-vivo" reports are available on the potential effects of thoracic stereotactic ablative radiotherapy (SABR) regimens to CIEDs functioning. The purpose of our study is therefore to evaluate the effects of SABR on CIEDs (pacemakers [PM] or implantable cardiac defibrillators [ICD]) in a cohort of patients affected by primary or metastatic lung lesions.
Materials and methods: We retrospectively collected all CIED-bearing patients undergoing SABR between 2007 and 2019 at our Institution. All CIEDs were interrogated before and after the SABR course to check for any malfunction. Prescription dose, beam energy and maximum dose (Dmax) to CIEDs were retrieved for each patient. Electrical records of the CIEDs were reviewed by the medical records.
Results: Thirty-four consecutive patients (24 with a PM and 10 with an ICD), who underwent 38 separate SABR courses, were included in the study. Eight patients (24%) were PM-dependent. Prescription dose of SABR ranged 26-60 Gy in 1-8 fractions, with a photon energy ranging 6-to-10 MV (76.3% and 23.7%, respectively) and a median Dmax to CIEDs of 0.17 Gy (range 0.04-1.97 Gy). Electrical parameters were stable in post-treatment device programming visits and no transient or persistent alteration of the CIED function was recorded in any patient. No inappropriate interventions were recorded in the 10 ICD-bearing patients during the treatment fractions.
Conclusions: Thoracic SABR proved to be safe for CIEDs when the dose is kept <2 Gy and the beam energy is ≤10 MV, irrespective of the pacing-dependency and of the CIED type.
Keywords: CIED; Cardiac device malfunction; Implantable cardioverter defibrillator; Pacemaker; SABR; Stereotactic radiotherapy.
Copyright © 2020 Elsevier B.V. All rights reserved.
Similar articles
-
A systematic review and meta-analysis on oncological radiotherapy in patients with a cardiac implantable electronic device: Prevalence and predictors of device malfunction in 3121 patients.Eur J Clin Invest. 2023 Jan;53(1):e13862. doi: 10.1111/eci.13862. Epub 2022 Sep 7. Eur J Clin Invest. 2023. PMID: 36004486 Free PMC article.
-
Radiotherapy-Induced Malfunction in Contemporary Cardiovascular Implantable Electronic Devices: Clinical Incidence and Predictors.JAMA Oncol. 2015 Aug;1(5):624-32. doi: 10.1001/jamaoncol.2015.1787. JAMA Oncol. 2015. PMID: 26181143
-
Radiotherapy-induced malfunctions of cardiac implantable electronic devices in cancer patients.Intern Emerg Med. 2020 Sep;15(6):967-973. doi: 10.1007/s11739-019-02240-y. Epub 2019 Dec 2. Intern Emerg Med. 2020. PMID: 31792775
-
Effect of X-ray dose rates higher than 8 Gy/min on the functioning of cardiac implantable electronic devices.J Radiat Res. 2020 May 22;61(3):419-425. doi: 10.1093/jrr/rraa016. J Radiat Res. 2020. PMID: 32253430 Free PMC article.
-
Current Clinical Practice in Patients With Cardiac Implantable Electronic Devices (CIED) Undergoing Radiotherapy (RT).Heart Lung Circ. 2022 Mar;31(3):327-340. doi: 10.1016/j.hlc.2021.10.020. Epub 2021 Nov 26. Heart Lung Circ. 2022. PMID: 34844904 Review.
Cited by
-
Impact of Radiotherapy on Malfunctions and Battery Life of Cardiac Implantable Electronic Devices in Cancer Patients.Cancers (Basel). 2023 Oct 2;15(19):4830. doi: 10.3390/cancers15194830. Cancers (Basel). 2023. PMID: 37835524 Free PMC article.
-
In-vitro effects of modern radiotherapy regimes on cardiac implanted electrical devices.Phys Imaging Radiat Oncol. 2025 Jul 2;35:100804. doi: 10.1016/j.phro.2025.100804. eCollection 2025 Jul. Phys Imaging Radiat Oncol. 2025. PMID: 40687307 Free PMC article.
-
A Review and Analysis of Managing Commonly Seen Implanted Devices for Patients Undergoing Radiation Therapy.Adv Radiat Oncol. 2021 Aug 6;6(4):100732. doi: 10.1016/j.adro.2021.100732. eCollection 2021 Jul-Aug. Adv Radiat Oncol. 2021. PMID: 34409216 Free PMC article. Review.
-
A case report of long-term successful stereotactic arrhythmia radioablation in a cardiac contractility modulation device carrier with giant left atrium, including a detailed dosimetric analysis.Front Cardiovasc Med. 2022 Aug 22;9:934686. doi: 10.3389/fcvm.2022.934686. eCollection 2022. Front Cardiovasc Med. 2022. PMID: 36072883 Free PMC article.
-
A systematic review and meta-analysis on oncological radiotherapy in patients with a cardiac implantable electronic device: Prevalence and predictors of device malfunction in 3121 patients.Eur J Clin Invest. 2023 Jan;53(1):e13862. doi: 10.1111/eci.13862. Epub 2022 Sep 7. Eur J Clin Invest. 2023. PMID: 36004486 Free PMC article.
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical