. 2021 May;25(3):213-221.
doi: 10.1016/j.canrad.2020.07.009. Epub 2021 Jan 2.

Total body irradiation using helical tomotherapy: Set-up experience and in-vivo dosimetric evaluation

Affiliations

Total body irradiation using helical tomotherapy: Set-up experience and in-vivo dosimetric evaluation

J-P Cleuziou et al. Cancer Radiother. 2021 May.

Abstract

Purpose: Helical Tomotherapy (HT) appears as a valuable technique for total body irradiation (TBI) to create highly homogeneous and conformal dose distributions with more precise repositioning than conventional TBI techniques. The aim of this work is to describe the technique implementation, including treatment preparation, planning and dosimetric monitoring of TBI delivered in our institution from October 2016 to March 2019.

Material and method: Prior to patient care, irradiation protocol was set up using physical phantoms. Gafchromic films were used to assess dose distribution homogeneity and evaluate imprecise patient positioning impact. Sixteen patients' irradiations with a prescribed dose of 12Gy were delivered in 6 fractions of 2Gy over 3 days. Pre-treatment quality assurance (QA) was performed for the verification of dose distributions at selected positions. In addition, in-vivo dosimetry was carried out using optically stimulated luminescence dosimeters (OSLD).

Results: Planning evaluation, as well as results of pre-treatment verifications, are presented. In-vivo dosimetry showed the strong consistency of OSLD measured doses. OSLD mean relative dose differences between measurement and calculation were respectively +0,96% and -2% for armpit and hands locations, suggesting better reliability for armpit OSLD positioning. Repercussion of both longitudinal and transversal positioning inaccuracies on phantoms is depicted up to 2cm shifts.

Conclusion: The full methodology to set up TBI protocol, as well as dosimetric evaluation and pre-treatment QA, were presented. Our investigations reveal strong correspondence between planned and delivered doses shedding light on the dose reliability of OSLD for HT based TBI in-vivo dosimetry.

Keywords: Helical tomotherapy; Irradiation corporelle totale; Tomothérapie hélicoïdale; Total body irradiation.

PubMed Disclaimer

MeSH terms

LinkOut - more resources